1 General information

1.1 Training

Self-instruction packet or training workshop and at least 1 installation

1.2 Required documents

Planning Guide
MAMMOMAT Balance Wiring Diagrams
Installation and Start-up Instructions
Maintenance protocol
Instructions for use
Service Instruction

1.3 Required tools, measurement and auxiliary devices

Oscilloscope
Oscilloscope probe 1:100 > 600Vdc
Digital multimeter
Non invasive kVp-meter (mammo)
Dosemeter (mammo)
Service PC with serial port (connection cable is supplied with the unit)
MAMMOMAT Balance SSW (Service Software Program is supplied with the unit)
SIB phantom or Normi 7 or 4 cm Plexi or customer phantom (supplied by customer)
Sensitometer / Densitometer
Film pack 18x24 and 24x30 if it occurs
Cleaning agents specified in the operating instructions
Protective conductor meter
Bathroom balance with tennis ball
Spray enamel according to the PG (to touch up chipped paint)
Standard installation tools
Plexiglas phantom 2cm, 2cm, 2cm, 1cm, 0.5cm for AEC calibration (supplied with the unit)

1.4 Required lubricants

All purpose grease PD2
Viscogen oil
MAMMOMAT Balance
Maintenance Instructions
SPB7-115.831.01
Page 5 of 22
Rev.01
05.04
1 General information
SPB7-115.891.01.01.02
SPB7-115.844.01.01.02
SPB7-115.814.01.01.02
SPB7-115.832.01.01.02
SPB7-115.629.01.01.02
SPB7-115.840.01.01.02