Model 1 1793A
Microwave Converter
LO Amplitude Range
+8 dBm to +13 dBm
$7
dBm to +13 dBm
0 dBm to
+5
dBm
1.
GENERAL INFORMATION
Frequency (GHz)
Conditions
2 to 18
18 to 26.5
18 to 26.5
Without Option 001, 011 or 021
Without Option 001, 011 or 021
With Option 001, 011
or
021
This Operating and Service Manual contains information about initial inspection, operation, performance
tests, adjustments, troubleshooting and repair of the HP Model 1 1793A Microwave Converter.
2 ,
Specifications
The HP 1 1793A Microwave Converter does not have "hard specifications" against which the instrument
can be tested for performance standards or limits. Instead, the Local Oscillator (LO) Requirements are
listed in Table 1. These requirements are the performance standards or limits against which the
instrument can be tested. The supplemental characteristics listed in Table
2
are not warranted
requirements but are typical characteristics included as additional information for the user.
Table 2. Supplemental Characteristics
RF Input Connector:
Precision 3.5 mm male
LO Input Connector:
Precision 3.5 mm male
IF Output Connector: Type-N female
Rear Panel Control Connector:
BNC female
Dimensions:
Height:
88 mm (3.5 in.)
Width:
425 mm (16.7 in.)
Depth:
528 mm (20.8 in.)
HP System II Module Size:
Height:
88 mm (3.5 in.)
Width:
1 Modular Width
Depth:
508 mm (20 in.)