System Care and Preventive Maintenance
CAUTION
1. Verify the operation of the black and white printer; freeze an image and perform a test
print. If necessary, clean the print head and test the printer again. If necessary, opti-
mize the printer settings according to the customer guidance. Refer to the OEM operat-
ing instructions for detailed information on optimizing printer settings.
2. Check other installed Siemens-approved OEM equipment. If necessary, clean the
OEM, and optimize the settings.
1.2.6

Concluding the Maintenance

1. Check system preset data, and restore from the disk if necessary.
2. Check the protective ground wire again. For detailed information, refer to
ARTD-002.731.17.01.02 (or newer).
3. Discuss the maintenance findings noted in the maintenance report with the customer.
Give the customer the completed maintenance report, and return a copy to your region-
al service office, as appropriate to your region.
NOTE
Siemens
Medical Solutions USA, Inc.
Verify only Siemens-approved devices are used.
Siemens is not responsible for damage or injury result-
ing from the use of non-Siemens approved devices. For
details, see the System Reference Manual, "Accessories
and Options".
Always retain a copy of the completed maintenance report for the
CSE's records.
US06-102.831.01.01.02