Recording Of Sw Option Keys, Tcp/Ip And Service Settings - GE Vivid 7 Service Manual

Hide thumbs Also See for Vivid 7:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FC091194, R
IRECTION
8-3-9

Recording of SW Option Keys, TCP/IP and Service Settings

NOTE:
Service Key is not required and should NOT be installed prior to Upgrade.
8-3-9-1
Record SW Option Keys
1.) Select CONFIG (F2) on the alphanumeric Keyboard on the operator panel.
2.) From the Operator Login window select adm from the Operator pull-down menu.
3.) Select LOG ON.
4.) From ADMIN tab, select the SYSTEM ADMIN.
5.) Record SW Options Keys from SW Options Key window.
Make sure that the SW Option Keys (passwords) have been recorded correctly. If the
WARNING
WARNING
key is incorrect, you will not be able to log on after the SW installation has been
completed. The password is case sensitive. Hyphens must also be recorded. There may
be more than one password.
Table 8-6
8 - 18
02
EVISION
Figure 8-1 System Admin
SW Options Keys
SW OPTIONS KEYS
Section 8-3 - SW Loading Procedure
V
7 / V
7 PRO S
IVID
IVID
ERVICE
M
ANUAL