Configure Internet Browser
Revision C
Equipment Overview: Service Utilities
7.
Type IP address, subnet mask and default gateway values matching the Unity
Network IX network domain for the ApexPro Telemetry Server. For example:
IP Address: 192.168.x.x
Netmask: 255.255.0.0
8.
Click OK.
9.
Click OK.
10. Click Yes to restart the service laptop.
Windows 2000
From the Windows taskbar, click Start > Settings > Network and Dial-up
1.
Connections.
2.
Right click Local Area Connection.
3.
Select Properties.
4.
Click Internet Protocol (TCP/IP).
5.
Click Properties.
6.
Record the current IP address and subnet mask so you can restore them later.
7.
Click Use the following IP address.
8.
Type IP address, subnet mask and default gateway values matching the Unity
Network IX network domain for the ApexPro Telemetry Server. For example:
IP Address: 192.168.x.x
Netmask: 255.255.0.0
9.
Click OK.
10. Click OK.
11. For Windows 2000, instead of restarting the service laptop, you can click Start
> Settings > Network and Dial-up Connections, right-click on Local Area
Connection, select Disable and then right-click again and select Enable.
The FE's Internet browser is configured to use GE specific configuration script to
access GE internal websites. The following steps must be performed to gain remote
access to ATS within the hospital intranet (ATS Webmin / ATS Desktop):
1.
Right-click on the Internet Explorer icon on the Desktop and select Properties.
2.
Select the Connections tab.
3.
Click LAN settings.
4.
Check Automatically detect settings.
5.
Uncheck Use automatic configuration script.
6.
Uncheck Use a proxy server.
7.
Click OK.
ApexPro Telemetry Server