5.2

Containment and transportation

Putting the contaminated equipment into exclusive shock and damage proof
container for transportation is recommended. It is recommended that
instruments are reprocessed as soon as possible and not later than 4
hours after usage.
5.3

Manual Cleaning and disinfection

Prepare
following
cleaning and disinfection:
a) Detergent: Cidezyme (Johnson & Johnson,
#2258) or another cleaning agent with
approved material compatibility for this
medical device
b) Disinfectant: Cidex OPA (Johnson & Johnson,
# 20391) or another disinfectant with
approved material compatibility for this
medical device
c) Two tanks, one for cleaning and one for
disinfection - optional:
1 additional tank for rinsing with deionized/tap water (sufficient
size for immersion of the immersible part of the probe at full length)
d) Soft, fluff free cloth or single use towel
e) Personal protective equipment (gloves, water repellent protective
skirt, face protection mask or protective glasses, see also
instructions of the manufacturer for the detergent and the
disinfectant)
items
before
- 17 -
Containment and
transportation
manual
Manual cleaning and
disinfection
Manual Cleaning
Rinsing after manual
Manual Disinfection
Rinsing after manual
disinfection
cleaning
Q1E-EP1432