Sanyo MLS-3020U Instruction Manual page 5

Labo autoclave
Table of Contents

Advertisement

WARNINGS AND CAUTIONS
7 Do not remodel the plug to avoid a fire hazard or injury to
*
presonnel by unusual action of the sterilizer.
Do not leave the power code on the sterilizer, if you leave it
that could cause a fire hazard.
The unit should be plugged into a dedicated electrical line
conforming to the following specifications.
MLS-3020U :
• 120Vunits : 120VAC(50/60Hz), 20A
• 208Vunits I 208VAC(50/60Hz), 10Aor greater
• 230Vunits : 230VAC(50/60Hz), 10Aor greater
MLS-2420U !
• 120Vunits : 120VAC(50/60Hz), 12.5A or greater
• 208Vunits : 208VAC(50/60Hz), 7.5Aor greater
• 230Vunits : 230VAC(50/60Hz), 7Aor greater
Failure to follow this instruction could result in a fire hazard
and injury personnel.
10 Never allow the unit to share an outlet with another electrical
appliance.
Do not use an ordinary table tap or an extension cord with
this sterilizer to avoid a fire hazard.
11 Do not put anything on the power supply cord, that could
cause a fire hazard.
Do not fix the power supply cord with pins such as staple or
nails, that could cause a fire hazard.
AC120V
AC208V
AC230V
4