7. Safety Precautions

 Never use the probe if the probe head, housing or cable are cracked
or damaged.
 When use this Probe (EUP-B712) for biopsy purpose,
use Needle Guide Bracket EZU-PA7C2 (Option) certainly.
 By examination of an early pregnancy the exposure time shall be as
short as possible. Start examination with acoustic output power set
to L (Low).
 The probe connector is not waterproof. Do not allow liquid to contact
the connector.
 Do not drop, hit or bend the probe.
 Use only water, detergents and disinfectants in the suppliers list.
 Under sterile condition, use appropriate protection for probe and
cable. Some Latex material may create allergic reactions.
 When the probe, the Magnetic sensor attachment and the Spacer for
EZU-RV2S are delivered, they are not cleaned, disinfected and
sterilized. Before using them, clean, disinfect and if necessary
sterilize them certainly.
 The acoustic lens of the probe is manufactured very thin and delicate
to get the high resolution. Therefore, in case of wiping out the
ultrasound jelly or cleaning the surface of the acoustic lens, please
use the soft cloth or tissue paper and handle with care.
WARNING
WARNING
CAUTION
CAUTION