Modules With Recorder, N-Frec And N-Fcrec - GE Healthcare B30 User's Reference Manual

Patient monitor
Hide thumbs Also See for Healthcare B30:
Table of Contents

Advertisement

Default alarm limits
EtCO
high 8%, low 3%
2
FiCO
high 3%, low Off
2
Carbon dioxide (CO
Measurement range:
Resolution:
Measurement rise time:
Accuracy:
Valid for respiration rate < 40 1/min at I:E ratio of 1:1. (Relative error is typically 10% for respiration rate 80
1/min at I:E ratio of 1:1.) The accuracy is specified in simulated ventilation. With higher respiration rates
and with varying ventilation methods the specifications may not be met.
Respiration rate
Breath detection:
Measurement range:
Accuracy:
Resolution:
NOTE: CO

Modules with recorder, N-FREC and N-FCREC

Power consumption
Recorder type:
Print resolution:
Paper width:
Traces:
Print speed:
1 Alarm limits and their adjustment range may vary depending on the mode used.
1
:
)
2
0 to 15 vol%
15 to 20 vol%
measurement is intended for patients who weigh over 5 kg (11 lb).
2
0 to 20 vol %
0.01%
< 300 ms with nominal flow
± (0.2 vol % + 2% of reading)
± (0.7 vol % + 2% of reading)
1% change in CO
level
2
4 to 80 breaths/min
±1 breath/min in the range 4 to 20 breath/min,
±5% in the range 20 to 80 breath/min
1 breath/min
3 W
Thermal array
vertical 8 dots/mm
horizontal 24 dots/mm
50 mm, printing width 48 mm
Selectable 1, 2 or 3 waveforms
1, 6.25, 12.5, 25 mm/s
System description
2-35