Clinical

This section answers questions regarding clinical issues for the system.
Resting ECG Report Format
Q: How do I change the way an ECG looks (format) when it prints out?
A: Follow these steps to change the ECG format:
1.
On the Main Menu, press F5 (System Configuration).
2.
Press F2 (Resting ECG).
3.
Press F4 (Page Down) two times.
4.
Select a Lead Sequence and Rhythm Leads.
5.
Press F6 (Save).
Editing
Q: Can you edit the interpretation at the device and then transmit the edited record to
the MUSE system as an unconfirmed record?
A: If you edit demographic information, only the record is still transmitted to the MUSE
system as an unconfirmed record.
If you edit the interpretation, the data is not saved unless the record is confirmed at
the device. The record is transmitted to the MUSE system as a confirmed record.

Navigating the User Interface

Q: How do I navigate from the Startup screen to the Main Menu?
A: You can configure the system in a number of different ways. Some of these
configuration choices determine the actions that you need to perform in order to
proceed from the power up display to the Main Menu.
2028451-183E
MAC™ 1600
Troubleshooting
55