Probes
Maximum probe temperature
Probe
M5Sc-D (Vivid S70 only)
3Sc-RS
6S-D
12S-D
9L-D
11L-D
L8-18i-D
C1-5-D
C1-6-D
C2-9-D
C3-10-D
IC5-9-D
2D (P2D)
6D (P6D)
6VT-D
6Tc-RS
9T-RS
10T-D (Vivid S70 only)
TM
AcuNav
10F
TM
AcuNav
8F
TM
SoundStar
3D 10F
TM
SoundStar
eco 10F
TM
SoundStar
eco 8F
NOTE:
13-8
Max Temp (°C)
(Simulated use)
42.0
41.1
41.3
41.5
41.6
41.8
41.9
41.1
41.4
38.6
41.1
41.9
35.7
34.8
40.8
41.7
41.6
41.9
37.0
37.8
37.0
37.0
37.8
Lens temperature measured under following conditions per
IEC 60601-2-37 Ed2:
1. Thermocouple is placed at the maximum temperature spot
of the lens. Ambient temperature at 23 °C ±3 °C.
Max Temp (°C)
(Still air)
42.2
37.1
39.9
33.1
37.6
38.0
36.5
40.5
39.5
37.3
36.8
35.4
35.5
28.1
36.0
37.9
33.6
42.5
34.9
33.6
34.9
34.9
33.6
–
Vivid S70 / S60
User Manual
BC092760-1EN 01