Chapter 5: Troubleshooting

Status

5.4.4.4 Sensor Housing Temperature Sensor Temperature to Resistance Curve

Temperature
15 °C
25 °C
30 °C
35 °C
40 °C
45 °C
50 °C
60 °C
5.5 Status
Access the Status screen from the main service menu. See "Accessing Service Mode" on page 4-1.
The Status screen displays status of all self tests the software runs continuously. These include: incubator
heater on (INCHTRON), warmer heater on (WRMHTRON), incubator/warmer heater off (I/WHTROFF),
humidifier heater on (RHHTR), remote monitoring data stream (RS-232LOOP), incubator fan on (FANON), and
incubator fan off (FANOFF). The current software revision of the options installed on this specific device
(Humidifier, Scale, SPO2, etc.) is shown.
If the RS-232 is not installed RS-232LOOP will display N/A. To test the circuit if the RS-232 is working, connect
the RS-232 to a working RS-232 compatible device or short pins 2 and 3 on the 9 pin connector.
5-22

FIGURE 5-2. Status Screen

Resistance
15.720 kOhm
10.000 kOhm
8.056 kOhm
6.530 kOhm
5.326 kOhm
4.369 kOhm
3.604 kOhm
2.491 kOhm
Service Manual