Syringe-Piping Combination - Hitachi L-2200 Instruction Manual

Optional syringe autosampler
Table of Contents

Advertisement

1.2 Syringe-Piping Combination

1.2 Syringe-Piping Combination
Syringe-
Piping
Syringe
Combina-
Capacity
tion Part
No.
0.1 mL*
0.5 mL
890-3511
1 mL
890-3512
2.5 mL
890-3513
5 mL
890-3514
* Furnished as a standard syringe kit for the L-2200 autosampler.
Five kinds of syringes are available as shown in Table 1-2.
For each syringe capacity, a sample loop for "needle - injection
valve connection" and a connecting coil for "injection valve -
syringe valve connection" are provided (see Figures 1-1 to 1-3).
Be sure to use any one of the syringe-piping combinations
indicated in Table 1-2.
To install an optional syringe onto the L-2200U autosampler,
remove the stainless pipe (P/N 890-3452, 0.1 mm ID × 0.36 m)
connected between the injection port and the injection valve port
3 on the L-2200U autosampler, and instead of it, connect the
stainless pipe (P/N 890-3451, 0.17 mm ID × 0.36 m) included in
the accessory box accompanying the L-2200 autosampler.
In an application where a considerably high flow rate is used
(e.g., a flow rate of 6 mL/mm or higher), replace the stainless
pipe connected between the injection port and the injection valve
port 3 with the stainless pipe (P/N 890-3473, 0.5 mm ID ×
0.36 m) furnished together with the optional syringe.
Table 1-2 Kinds of Syringes
Sample Loop
(For connection between needle
and port 6 of injection valve)
Syringe
Part No.
Part No.
890-3228
890-3462
080-1064
890-3481
890-3466
080-1592
890-3481
890-3468
080-1092
890-3481
890-3470
080-1593
890-3464
Connecting Coil
(For connection between port 5
of injection valve and port 3 of
syringe valve)
Dimensions
Part No.
(Capacity)
0.5 mm ID × 0.7 m
890-3455
(0.14 mL)
0.5 mm ID × 0.7 m
890-3467
[Loop 1]
(0.14 mL)
0.8 mm ID × 0.8 m
[Loop 2]
(0.4 mL)
0.5 mm ID × 0.7 m
890-3469
[Loop 1]
(0.14 mL)
1.0 mm ID × 1.2 m
[Loop 2]
(0.9 mL)
0.5 mm ID × 0.7 m
890-3471
[Loop 1]
(0.14 mL)
1.0 mm ID × 3.1 m
[Loop 2]
(2.4 mL)
1.0 mm ID × 6.4 m
890-3465
(5.0 mL)
1 - 2
Dimensions
(Capacity)
0.25 mm ID × 1.0 m
(0.05 mL)
0.8 mm ID × 1.0 m
(0.5 mL)
1.0 mm ID × 1.3 m
(1.0 mL)
1.0 mm ID × 3.2 m
(2.5 mL)
1.0 mm ID × 6.4 m
(5.0 mL)

Advertisement

Table of Contents
loading

This manual is also suitable for:

L-2200u

Table of Contents