Replacement Parts; Consumables - Hitachi L-2200 Instruction Manual

Optional syringe autosampler
Table of Contents

Advertisement

5.1 Consumables

Part No.
Part Name
080-1064
Syringe (0.5 mL)
080-1592
Syringe (1 mL)
080-1092
Syringe (2.5 mL)
080-1593
Syringe (5 mL)
890-3481
Sample loop
(0.5 mm ID × 0.7 m)
890-3464
Sample loop
(1.0 mm ID × 6.4 m)

5. REPLACEMENT PARTS

Listed below are the parts to be prepared at hand as
consumables for replacement. The useful-year period of each
part is based on a situation where the instrument is operated for
250 days per year, for 8 hours per day. The standard lifetime of
each part indicates the maximum allowable number of plunger
reciprocations in pure water feeding. Since this value varies
significantly depending to analytical conditions, keep an
adequate quantity of consumable parts at hand in consideration
of actual operations to be performed.
Location
Syringe
valve
Syringe
valve
Syringe
valve
Syringe
valve
Flow
path
Flow
path
NOTICE: Keep the consumable parts at hand and replace
them periodically as specified. If they are not
replaced periodically, it may result in performance
degradation or occurrence of a failure.
5 - 1
Useful
Standard Lifetime
Years
2 years
Plunger reciprocation:
40,000 times
1 year
Plunger reciprocation:
20,000 times
1 year
Plunger reciprocation:
20,000 times
1 year
Plunger reciprocation:
20,000 times
5 years
Frequency of analysis:
60,000 times
3 years
Frequency of analysis:
36,000 times
5.1
Remarks
For 0.5 mL
syringe
For 1 mL
syringe
For 2.5 mL
syringe
For 5 mL
syringe
For 0.5 mL/
1 mL/2.5 mL
syringe
For 5 mL
syringe
(See 1.2)

Advertisement

Table of Contents
loading

This manual is also suitable for:

L-2200u

Table of Contents