Installation; Unpacking; Setup; Wash Count Parameters - Hitachi L-2200 Instruction Manual

Optional syringe autosampler
Table of Contents

Advertisement

2.1 Unpacking

2.2 Setup

2.2.1 How to Set
Syringe Capacity
and Wash Count
Parameters

2. INSTALLATION

NOTICE: It is not allowed for the user to perform installation
at the time of delivery of the instrument.
For ensuring safety and high accuracy of the
instrument, installation shall be carried out by or
under supervision of qualified service personnel of
the manufacturer or its authorized service agent.
Before installation of the instrument, the customer
is requested to make necessary preparation
referring to this manual. If relocation of the
instrument becomes necessary after initial
installation (delivery), please notify your local
Hitachi High-Technologies Corporation sales
representative or service office of Hitachi High-
Technologies Corporation sales representative.
Install this optional syringe in the inside of the L-2200/2200U
autosampler in the manner instructed below.
(1)
Referring to the L-2200 Autosampler Instruction Manual -
4.3.2 (3) "Setting Operating Conditions" - (a) "Sequential
Parameters" (L-2200U Autosampler Instruction Manual -
4.3.2 (3) "Setting Operation Conditions" - (a) "Sequential
Parameters"), enter a code corresponding to the capacity
of the optional syringe to be used (SYRINGE VOL).
(2)
Referring to the L-2200 Autosampler Instruction Manual -
4.3.2 (3) "Setting Operating Conditions" - (c) "WASH
PARAM" (L-2200U Autosampler Instruction Manual - 4.3.2
(3) "Setting Operation Conditions" - (c) "WASH PARAM"),
specify "1" for INJ WASH STROKES that is a wash
condition setting parameter in the sequential program to
be carried out.
2 - 1