Health O Meter 498KL User Instructions

Health O Meter 498KL User Instructions

Remote display digital scale
Hide thumbs Also See for 498KL:

Advertisement

Quick Links

Model 498KL / 498KL-USB
Remote Display Digital Scale
User Instructions
E112697
UL 60601-1
CSA C22.2 No. 601.1-M90
Para ver las instrucciones de uso en español visita www.homscales.com
Pour les instructions de l'utilisateur dans la visite française www.homscales.com
PELSTAR, LLC
9500 West 55th St. McCook, IL 60525-7110 USA
www.homscales.com
®
Health o meter
Professional 498KL
© Pelstar, LLC 2021
P/N UM498KL Rev20210930_
1

Advertisement

Table of Contents
loading

Summary of Contents for Health O Meter 498KL

  • Page 1 Model 498KL / 498KL-USB Remote Display Digital Scale User Instructions E112697 UL 60601-1 CSA C22.2 No. 601.1-M90 Para ver las instrucciones de uso en español visita www.homscales.com Pour les instructions de l'utilisateur dans la visite française www.homscales.com PELSTAR, LLC 9500 West 55th St. McCook, IL 60525-7110 USA www.homscales.com...
  • Page 2: Table Of Contents

    Professional 498KL ® Health o meter Professional 498KL-USB Thank you for purchasing this Health o meter® Professional product. Please read this manual carefully, and keep it for easy reference or training. TABLE OF CONTENTS Caution and Warning....................3 Specifications for This Scale ..................4 Certifications / Disposal .....................
  • Page 3: Caution And Warning

    When operating the scale with the AC adapter:  Operate this scale exclusively with the AC adapter provided by Health o meter Professional. Use of an unspecified adapter will void the warranty, and can pose a serious safety hazard.
  • Page 4: Specifications For This Scale

    SPECIFICATIONS FOR THIS SCALE General The model 498KL remote display digital scale uses highly sophisticated microprocessor technology. This precision scale is designed to provide accurate, reliable and repeatable weight measurements. In addition, each scale is designed to provide the user with features that make the weighing process simple, fast, and convenient.
  • Page 5: Certifications / Disposal

    CERTIFICATIONS / DISPOSAL Certification Descriptions Pelstar, LLC has been officially certified as the manufacturer of Health o meter® Professional medical devices. The Pelstar, LLC quality assurance system covers the development, production, sales and service of medical scales and measuring systems.
  • Page 6: Assembly Instructions 498Kl Scale

    498KL ASSEMBLY INSTRUCTIONS Before Assembly Each 498KL remote display digital scale is shipped disassembled in one carton. Carefully inspect the carton for shipping damage before unpacking. If damage is found, contact your shipper or a Health o meter® Professional representative immediately at 1-800-815-6615.
  • Page 7: Display Bracket Assembly Instructions

    DISPLAY BRACKET ASSEMBLY INSTRUCTIONS Display Bracket Assembly Note: Do not use display bracket for wall mount. Display bracket is for surface use only. For wall mounting, mount display assembly directly to the wall, see page 8. 1. Carefully remove the scale platform with the display assembly from the carton, and place the scale platform and display assembly on a flat, level, and dry surface.
  • Page 8: Display Wall Mount Instructions

    DISPLAY WALL MOUNT INSTRUCTIONS Caution: When wall mounting the display assembly, ensure that the mounting location is free of electrical lines, gas pipes, and other safety hazards. Ensure that wall is of construction that can support the display assembly. Note: Remote head can be mounted to a wall up to 6 feet (1.83 m) from the base or it can be mounted on the display bracket (included) for placement on a table up to 6 feet (1.83 m) from the base.
  • Page 9: Powering The Scale

    1. With a Phillips head screwdriver remove the battery cover on the side of the display assembly by unscrewing the (2) battery cover screws (see drawing below). 2. Replace or install 6 new AA batteries. Health o meter Professional recommends using alkaline or lithium batteries, rated for 130°F (54°C) operation. Do not use zinc-carbon batteries.
  • Page 10: Set Up Instructions

    SET UP INSTRUCTIONS Auto-Off Function As an energy saving feature, the scale will automatically turn off after 2 minutes of inactivity. By default, this function is enabled. Disable Auto-Off 1. While the scale is turned off, press and hold the HOLD/RELEASE button. 2.
  • Page 11: Operating Instructions

    OPERATING INSTRUCTIONS Keypad Function Description ON/OFF Turns scale on and off. ZERO Tares the scale prior to a weighing. Holds the weight value of the patient on the display until the button is HOLD/RELEASE pressed again to clear the value. Prompts entry of height data to calculate patient’s Body Mass Index (BMI).
  • Page 12 OPERATING INSTRUCTIONS (CONT) Weighing a Patient Note: When this scale is turned on, it will always default to the settings and units last used (LB or KG). 1. With all weight off the scale, press the ON/OFF button to turn on the scale. 2.
  • Page 13 OPERATING INSTRUCTIONS (CONT) BMI Function Body Mass Index (BMI) is a ratio between a person’s weight and height and interpreted as a BMI score. This BMI score is then used as a tool to categorize a patient as underweight, normal, overweight, obese or morbidly obese.
  • Page 14: Transport

    TRANSPORT This scale can be easily placed into an optional carrying case for storage or transport from one location to another. This scale does not include the carrying case item # 64771, which may be acquired separately. To prepare the scale for transport, follow these steps: 1.
  • Page 15: Maintenance

    4. Do not pour or spray fluids directly on the scale. Replacement Parts Parts may be available if a piece of the scale needs to be replaced. Contact Health o meter Professional Scales Customer Service at 1-800-815-6615 to inquire about the availability of these replacement parts.
  • Page 16: Calibration

    This scale has been factory calibrated, and does not require calibration prior to use. If required, the scale can be calibrated. Calibration of your 498KL scale is performed using kilograms (KG) or pounds (LB), according to the unit of measure used upon entering into the calibration path.
  • Page 17: Troubleshooting

    TROUBLESHOOTING Troubleshooting Before contacting service personnel, refer to the following instructions to check and correct any failures. Symptom Possible Cause Corrective Action Scale does not turn on 1. Dead battery 1. Replace batteries 2. Faulty electrical outlet 2. Use a different outlet 3.
  • Page 18: Warranty

    What does the Warranty Cover? This Health o meter® Professional scale is warranted from date of purchase against defects of materials or in workmanship for a period of two (2) years. If product fails to function properly, return the product, freight prepaid and properly packed to Pelstar, LLC (see “To Get Warranty Service”, below, for instructions).