Product Lnstallation - Haier HCP-80 Operating Instructions Manual

Co2 incubator
Table of Contents

Advertisement

Product Installation
Product unpacking and handling
• Use a suitable unpacking tool such as scissors or wallpaper knife to cut the packing strap, Lift the
outer box vertically and move it horizontally; after that, check that if the accessories are complete.
Caution Avoid collision and scratches on the exterior of the incubator. For the sake of safety,
please have at least two adults for unpacking
• Since the incubator body is heavy, please choose suitable mechanical handling tools.
For manual handling, be sure to have four adults (at least two) to lift the incubator
Caution
vertically with hands holding the bottom of the incubator, and then move it horizontally
to the position where the incubator is to be stored. Do not apply pressure on the outer
door of the incubator during handling. The door body may be damaged.
Installation environment
• Ambient temperature: 18 °C ~ 32 °C;
• Ambient humidity: ≤ 80% Rh;
• Atmospheric pressure: 86kPa~106kPa;
• Indoor use;
• Installation environment must be free from dust, vibration and corrosion gas;
• Avoid direct sunlight or sources of cold and heat;
• Make sure no corrosive substances or high concentration of dust around the incubator;
• Power voltage: Within rated voltage±10%;
• Voltage frequency: 50Hz.
• Altitude of the working position of the box:below 2000m.
• Pollution Degree:2.
• The transient state is Category II facility (overvoltage category).
• The safety of any system incorporating the equipment is the responsibility of the assembler of the
system.
• The operation of the incubator requires CO
Warning
harmful to health. Therefore, the installation position shall have good ventilation
condition so that the exhaust gas from the rear wall can be quickly removed.
• The incubator shall not be placed in a place without a ventilation system. If there are
several incubators in the same room or the incubator is placed on the bottom layer of
the laboratory, an auxiliary ventilation system is recommended.
The expected performance of the incubator cannot be guaranteed when it is not used
Caution
in the specifi ed environment.
Spacing
•The incubator must have a space from the adjacent wall/object, as shown in the fi gure below.
Do not block the pressure compensation hole on the rear wall. Do not place the
Caution
incubator in any position where it is diffi cult to operate the disconnect device.
gas flow. Excessive CO
2
7
can be
2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hcp-168Hcp-258

Table of Contents