Philips ReCare 7.0 U User Manual page 58

Table of Contents

Advertisement

• Sensor test
- The bike has two sensors, one angle sensor for speed/velocity
measurements located on the brake, and one reed switch that
measures crank rotation which we use to determine crank
position.
- MW will show: ANGLE 0 REED 0
When sensors operate correctly: rotate the crank and the Angle
reading will show pedal RPM measurement and the Reed will
change from 0 to 1 once per pedal revolution.
• Crank position cali
- Software calibration to set the position of the right pedal at 12
o'clock. Set right pedal to 6 o'clock position then press start.
- Rotate the right pedal clockwise until the console beeps.
• Watts calibration (Factory use only)
• Unit type
- Selects Upright bike (7.0 U).
Error messages
EEPROM error: Solution for this is to replace the console (Note: this is
the only error message)
Troubleshooting
Below are common problems and basic checks to solve them. If these
tips do not solve your problem then call your local distributor for
service)
No power
• Make sure the A.C. outlet has power (90~240VAC), the line cord
plugged in securely and the power switch is on.
• Check the fuse in the Input module (located between the power
switch and line cord input).
• Make sure all connectors in back of the console are securely seated
in place.
58