Multi gas incubator
Table of Contents

Advertisement

CAUTION
Perform H
O
decontamination with the chamber attachments arranged as specified by Sanyo.
2
2
Arranging them in a different way may result in insufficient decontamination.
CAUTION
Wear rubber gloves when handling the H
in inflammation of the skin.
CAUTION
After H
O
decontamination has been completed, residual H
2
2
chamber, the H
O
Generator, and the bottom of the duct. Wearing protective glasses and rubber
2
2
gloves, wipe it off with a non-woven cloth. Failure to do so may result in a deficient culture.
 
reagent. Direct contact with the H
O
2
2
50
Get other manuals https://www.bkmanuals.com
O
2
2
O
solution will remain on the bottom of the
2
2
reagent may result