Example
Enter the value shown below if the displayed chamber CO
4.5%.
●CO
density span: Enter 045. 
2
 
 
T e m p / C O 2 / O 2
1
 
 
3
 
T e m p
 
4
C O 2
 
5
C O 2
 
O 2
 
 
6
O 2
 
Example
Enter the value shown below if the displayed chamber O
5.5%.
●O
density span: Enter 055. 
2
 
 
1
T e m p / C O 2 / O 2
 
 
3
 
T e m p
 
4
C O 2
 
5
C O 2
 
 
O 2
 
O 2
6
 
4. Press the MENU Key to display the menu, select OK, and press the ENTER Key.
5. After the Select Tools Display is displayed, press MENU Key to display the menu, select Cancel, and
press the ENTER Key to return to the Top Display.
Note:
It is important to accurately measure the temperature inside the Incubator when calibrating the
temperature. Particularly, the temperature gauge used must have an accuracy of JIS 0.5 Class or better.
The temperature must be measured at several points.
The temperature setting can not change by more than ±1.0
exceeded, an error buzzer will sound, the input data will be ignored, and the Incubator will return to the
display mode. If it is necessary to change the temperature by more than 1.0
several stages over a period of time.
C a l i b r a t i o n
o
S p a n
0 . 0
C
Z e r o
S p a n
4 . 5
%
Z e r o
S p a n
0 . 0 %
C a l i b r a t i o n
o
S p a n
0 . 0
C
Z e r o
S p a n
0 . 0 %
Z e r o
S p a n
5 . 5
%
Get other manuals https://www.bkmanuals.com
density is 5.0% but the actual CO
2
P V
V o l t
o
3 7 . 0
C
3 . 9 1
5 . 0 %
3 .
1 6
5 . 0 %
5 8
density is 5.0% but the actual O
2
P V
V o l t
o
3 7 . 0
C
3 . 9 1
5 . 0 %
3 .
1 6
5 . 0 %
5 8
o
C during calibration. If these values are
59
density is
2
0 . 5 ° C
V
5 0 0
V
7 0 0
5 0 0
m V
5 0 0
density is
2
0 . 5 ° C
V
5 0 0
V
7 0 0
5 0 0
m V
5 0 0
o
C, perform calibration in