Installing the WLAN Expansion
Press the power switch into the break-out. See
Wiring of the WLAN Power Supply Power Cable - Power Switch
Lay the cable from the WLAN power supply through the round opening
(4/Fig. 13 / p. 21)
Connect the cable with the designation M15 L on the WLAN power supply, terminal L
(5/Fig. 13 / p.
Connect the cable with the designation M15 N on the WLAN power supply, terminal N
(5/Fig. 13 / p.
Plug in the cable with the designation S15.1A on the power switch
Plug in the cable with the designation S15.2A on the power switch
Wiring the WLAN Power Switch with the Power Outlet Strip
Lay the cable from the power outlet strip up to the WLAN power switch.
Plug in the cable with the designation S15.1 on the power switch
Plug in the cable with the designation S15.2 on the power switch
Plug in the three-prong connector at the power outlet strip.
Wiring the Network Cable
Plug in the new network cable on the RJ45 socket on the WLAN Client
(9/Fig. 13 / p. 21)
Plug in the other end of the network cable at the next free socket on the network switch.
Secure the cable at suitable locations using cable ties.
Labels
Attach one of the labels with the warning symbol for non-ionizing radiation and a refer-
ence symbol to to the Operator Manual along side each other next to the WLAN power
switch.
Attach the 2nd label with the warning symbol for non-ionizing radiation and the 2nd ref-
erence symbol to the Operator Manual on the top console of the monitor carriage,
directly below the cable outlet for the mouse (= location of the signal source below the
console, centered to the signal antennas of the WLAN Client).
Siemens
Medical Solutions
Installation location of the switch: 0 label = top; 1 label = bottom.
up to the inside of the top compartment to the WLAN power switch.
21).
21).
NOTE
The loose ground wire covered with heat-shrink tubing that
is from the power cable does not have to be connected at
both ends.
NOTE
The loose ground wire from the power cable (switch side)
does not have to be connected. A ground wire connection to
the monitor carriage is already connected.
and lay it to the side to the existing network switch.
DRAFT 22.04.2008
SP00-000.814.09.01.02
11.07
(Fig. 14 / p.
22).
Page 23 of 78
CS PS SP
23
(1/Fig. 14 / p.
22).
(2/Fig. 14 / p.
22).
(3/Fig. 14 / p.
22).
(4/Fig. 14 / p.
22).