HP 3324A Operating And Programming Manual

Synthesized function/sweep generator
Hide thumbs Also See for 3324A:
Table of Contents

Advertisement

Quick Links

HP 3324A
Synthesized
Function/Sweep Generator
Operating and Programming Manual
HEWLETT
PACKARD

Advertisement

Table of Contents
loading

Summary of Contents for HP 3324A

  • Page 1 HP 3324A Synthesized Function/Sweep Generator Operating and Programming Manual HEWLETT PACKARD...
  • Page 3 Operating and Programming Manual HP 3324A Synthesized Function/Sweep Generator SERIAL NUMBERS This manual applies to instruments with serial number 2836GOOlll and higher. Any change made in instruments having serial numbers higher than the above numbers will be found in a “Manual Changes”...
  • Page 4 Notice Subject Matter Notice The information in this document is subject to change without notice. Hewlett-Packard makes no warranty of any kind with regard to this printed material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of...
  • Page 5 Printing History New editions are complete revisions of the manual. Update packages, which are issued between editions, contain additional and replacement pages to be merged into the manual by the customer. The date on the title page and back cover of the manual only changes when a new edition is published.
  • Page 6 Preface Introduction This manual provides the information required when either operating the HP 3324A from the front panel or programming it from a remote controller. What is in this Chapter 1: Instrument Introduction Gives a brief description of the manual? instrument and its features.
  • Page 7 Chapter 12: Performance Tests Gives the procedures to verify the in-spec performance of the BP 3324A. Specifications Appendix A: Lists the HP 3324A specifications. Appendix B: Options and Accessories Gives details of all the options and accessories available for the HP 3324A.
  • Page 8 During the warranty period, HP will, at its option, either repair or replace products which prove to be defective. For warranty service or repair, this product must be returned to a service facility designated by HP.
  • Page 9 The remedies provided herein are Buyer’s sole and Exclusive Remedies exclusive remedies. HP shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any other legal theory. Assistance Product maintenance agreements and other customer assistance agreements are available for Hewlett-Packard Hewlett-Packard Sales and Service Office.
  • Page 11: Table Of Contents

    Contents Instrument Introduction Introduction ... . . l - l Features ... . . l - l Operating Concept ..Local Mode .
  • Page 12 HP-IB Connector (16) ..Main Signal Output Connector (17) . . Phase Cal In (18) ..Phase Cal Out (19) ..
  • Page 13 ... . . Powering Up ... . . The HP 3324A as a Function Generator . The HP 3324A as a Sweep Generator . . 5. Parameters Introduction .
  • Page 14 Setting the Default Parameters ..Choosing the Number of Stores/Intervals Changing the HP-IB Address ..Choosing HP 3325A Compatibility . . . Performing a Self- test ..
  • Page 15 Omitting the Mnemonic ..9-11 . . . Synchronizing after a Syntax Error 9-11 Setting the HP-IB Address ..Status Reporting Introduction ... . . Status Byte .
  • Page 16 Performance Tests 12-1 Introduction ... . . 12-1 Equipment Required ..12-1 Test Record ... . . 12-2 Test Failure .
  • Page 17 Amplitude Accuracy at Frequencies upto 12-18 ..100 KHz Procedure 12-25 High-Voltage Output (Option 002) . . Amplitude Flatness (Frequencies above 12-27 100 KHz) ... . 12-29 High-Voltage Output (Option 002) .
  • Page 18 Auxiliary Outputs . . . Auxiliary Inputs (Supplementary) ..HP-IB Control (Supplementary) Option 001 High-Stability Frequency Reference ... . . A-10 A-10 Option 002 High-Voltage Output ..
  • Page 19 Line Voltage Selection ..Power Cable ... . . HP-IB Connector ... c - 7 c - 7 HP-IB Logic Levels .
  • Page 20 ... . . Syntax Drawing Rules ..HP-IB Command List ..AC; Amplitude Calibration .
  • Page 21 E-28 P H ; P h a s e ....E-29 ... RE; Recall State RF; Rear or Front Signal Output ..E-30 E-31 RTT;...
  • Page 22 Eastern USA ... . Midwestern USA ..Southern USA ..... . Western USA .
  • Page 23: Instrument Introduction

    Instrument Introduction The III’ 3324A Synthesized Function/Sweep Generator Introduction a multi-task generator, which can be used as: a reference source Produces a sinewave of a specified frcqucncy, amplitude, DC offset and phase. a function generator Produces various waveforms at a specified frequency, amplitude, DC offset and phase.
  • Page 24: Operating Concept

    Operating Concept The HP 3324A has been designed so that it can be used in either of two modes: a. Local Mode: Using the front panel to initiate operations. b. Remote Mode: Using an external computer as the controller. Local Mode...
  • Page 25: About This Manual

    About this Manual This manual is basically divided into two main parts: A part which describes the local control of the instrument (chapters 2 to 8). A part which describes the remote control of the instrument (chapters 9 to 11). The examples in the local control part are shown in such a way that on the left side of the page the keystrokes are shown, while the right side gives an explanation of...
  • Page 26: About The Note

    About the Note The NOTE sign is used to show where additional Note information beneficial to the reader can be found. This information refers to specific situations or operations. Instrument Introduction...
  • Page 27 - 0 P C...
  • Page 29: Introduction

    Front and Rear Panel Overview Introduction This chapter contains basic information about the front and rear panels of the HP 3324A. The descriptions given should be read in conjunction with the foldouts at the front of this chapter. Front Panel The front panel can be divided into 15 main component areas, each of which is described here.
  • Page 30: Select/Modify (4)

    SELECT/MODIFY (4) These keys are used to alter data and to move the cursor within menus or the display. The function of the keys depends on the current menu or value displayed. For more information see Chapter 3 Instrument Operation. The PARAMeter keys allow you to view, enter or change the frequency, amplitude, DC offset and phase of the...
  • Page 31: Main Signal Output Connector (10)

    If (Shift is pressed “S” is shown in the display until the next key is pressed. This key returns the HP 3324A from remote control to front-panel control unless local lockout has been programmed. The Iut;l) key accesses the Utilities menu, which contains...
  • Page 32: Hp-Ib Connector

    The rear panel can be divided into 13 main component areas, each of which are described here. HP-IB Connector (16) The HP-IB connector is used to connect the HP 3324A onto the Hewlett-Packard Interface Bus so that it can be used in remote mode and accessed by a controller.
  • Page 33: Ac Line (21)

    30 MHz oscillator. If an external reference is present “EXT-REF” is shown in the display. Ref Out (23) A 1 MIIz signal from the IIP 3324A reference circuits is available here. Aux 0 dBm (24) A squarewave signal is available at this output for frequencies between 21 MIIz and 60 MHz.
  • Page 35: Instrument Operation

    Instrument Operation Introduction This chapter describes the operating principles of the HP 3324R. It describes the powering on procedure, the various outputs available and how to use both the DATA keys and the SELECT/MODIFY keys. Before Power-on Caution Before connecting the HP 3324A...
  • Page 36: Turn-On State

    HP Service Center for repair. Turn-on State The initial state of the HP 3324A depends on if the instrument is in the HP 3325A compatibility mode or not. The mode can be set and seen in the Utilities menu (see Chapter 8).
  • Page 37 In the HP 3325A compatible mode all stores are deleted, and a default setting becomes the new state. Table 3-l shows the power-on state for the HP 3325A compatible mode. The main signal output is enabled at power-on (LED to the right of the (SignalG&ZKX) key is lit).
  • Page 38: Input/Output Signals

    Input/Output The HP 3324A has a total of 13 BNC connectors, located on both the front and rear panels. The use of Signals each connector is explained in the following sections. Caution A maximum of 42 V can be applied as an external voltage to any of BNC connectors because of the HP 3324A’s floating ground.
  • Page 39: Synchronization Outputs

    Phase Calibration This input must be connected to the Phase Cal Output of another HP 3324A. One of the instruments must Input have option 003 installed and the other option 004. The two instruments can then be phase calibrated. Use the...
  • Page 40: Phase Calibration Output

    Reference Input oscillators). Phase-locking to an external frequency reference transfers the external reference’s frequency accuracy and aging rate to the HP 3324A. The external reference controls the standard 30 MHz internal reference oscillator frequency. The external reference level must be from 0 dBm to +20 dBm (50R), and the frequency...
  • Page 41: Frequency Reference Output

    The Ref Out 1 MHz connector supplies a 1 MHz squarewave derived from the frequency reference of the output HP 3324A. The squarewave has a level greater than 0 an can be used to phase-lock an analyzer or other instrumentation to the frequency reference of the HP 3324A.
  • Page 42: X-Drive

    X-drive The X-drive output signal is only present during a sweep operation, as follows: Single sweep: 0 V at the start of the sweep, increasing linearily to > +lO V at the sweep stop. The voltage remains at this level until the reset prior the start of another sweep.
  • Page 43: Aux 0 Dbm 21-60 Mhz

    Aux 0 dBm 21-60 MHz The rear-panel Aux 0 dBm 21 - 60 MHz connector supplies a squarewave signal for frequencies between 21 and 60 MHz. The amplitude of the signal is 0 dBm. To obtain a 21 - 60 MHz frequency, the auxiliary TTL clock waveform (“AUX”) must be selected.
  • Page 44: Data Keys

    As on calculators the (BackSpace) key clears the digit to the left of the cursor. If an entered value exceeds the range limits, the HP 3324A ignores the entered value, briefly displays the error, and then puts the instrument into the edit mode.
  • Page 45: Select/Modify Keys

    SELECT/MODIFY The SELECT/MODIFY keys are used in conjunction with most of the other keys and the display. The four Keys arrow keys are referred as the a, a, a and a keys in this manual. Select The select mode is used to move through the menus to a certain item/point, which can then be activated with the (Select) key.
  • Page 46 There are two ways to change a digit in the modify mode: the numeric keypad. There is no need to press [Enter). Use the Ilncr)/m keys to increase/decrease the digit by 1. Holding the key in will increase/decrease the value at a faster rate. Note that, except on the most significant digit, increasing from 9 to 10 causes the 1 to be carried to the next digit position, and decreasing from 0 to 9 will cause the preceeding digit to be...
  • Page 47 Example Consider a frequency of 125 KHz. You wish to see what happens when you increase the frequency to 225 KHz and back again in steps of 10 KHz. Result The current frequency value is displayed is set. A frequency of 125 KHz entered modify mode is and the cursor is placed under...
  • Page 48: Example

    Example Consider an amplitude of 4.075 V(p-p). You wish to see what happens when you increase the amplitude to 4.125 Keystrokes The current amplitude is shown in the display. The amplitude is changed to 4.075 V(p-p). The edit mode is activated. The cursor moves under the “5”.
  • Page 49: Saving/Recalling Instrument States

    (St,: xx int.: XX) command contained in the Utilities menu (see Chapter 8). At power-on, if the HP 3324A is in the HP 3325A compatible mode, all the stores will be deleted. If the instrument is in the non-compatible mode, the stores will contain the same information as previous to power-off.
  • Page 50: Example

    Example Consider saving an instrument state in the 4th register. Keystrokes The present instrument setting is saved in register 4. operating Recalling Instrument state is recalled from the internal memory, when the m key, a digit and then (Enter) are pressed. States digit specifies the memory location from which the operating state is to be recalled.
  • Page 51: Getting Started

    Introduction This chapter gives a very brief description of how to get started with the HP 3324A. This includes how to power up, an example of setting the parameters for a standard function generator, and an example of setting up a sweep.
  • Page 52: The Hp 3324A As A Function Generator

    (LED to the right of the I-j switch is lit). The HP 3324A as The following is an example of how to set up the HP 3324A as a standard function generator. It shows how a Function to choose a waveform, set the frequency, amplitude and Generator offset of the output signal, and how to enable the signal.
  • Page 53 Keystrokes Result The triangle is selected as the signal waveform Note If the error “freq. out range” is shown here, do steps 3 and 4 before repeating steps 1 and 2. This is because the frequency is not valid for the waveform chosen. Keystrokes Result The parameter menu is shown in the display, along with the...
  • Page 54 Keystrokes Result The LED to the left, of the key lights, to show that the main signal output is enabled. This example has shown how to set up the HP 3324A as a function generator. 4-4 Getting Started...
  • Page 55: The Hp 3324A As A Sweep Generator

    The HP 3324A as a The following is an example of how to set up the 3324A as a sweep generator. The example follows Sweep Generator on from the one above. It shows how to set up three sweep intervals, and to cause a continuous sweep to be performed using these intervals.
  • Page 56 Note The sequence does not have to follow in series, that is 1, 2, 3, 1, 2. You can program the intervals in any sequence that you require, for example, 2, 3, 1, 3, 1, 2 and so on. Now, each of the intervals must be programmed separately.
  • Page 57 Keystrokes The stop frequency for the 2nd interval is changed to 10 1 5 . The sweep time for the 2nd interval is changed to 2 s. The cursor is now on “SWEEP-TIME”. The 2nd interval is set as a logarithmic sweep. The cursor is now on “LOG”...
  • Page 59: Parameters Introduction

    Parameters Introduction The Parameter keys (Freg), enable you to display the frequency, amplitude, DC offset and phase values, respectively, of your output signal. Figure 5-1. The Parameter Keys When one of the keys is pressed, the current value is displayed along with the parameter key functions and the current waveform.
  • Page 60: Entering A New Value

    K H z RMPL OFFS Figure 5-2. The Parameter Display One of the other parameters can be viewed by either pressing the parameter key, or using the 0J or a keys to move the cursor up or down the Parameter menu. Putting the cursor (so that the parameter blinks) on the amplitude, on “OFFS”...
  • Page 61: Modifying A Parameter

    Modifying A parameter can be simultaneously updated using the modify keys. This is done by activating the parameter Parameter that you wish to change, and then pressing the yellow “M” is shown in the display, to indicate that you are in the modify mode.
  • Page 62: Modify Example

    Modify Example Consider a frequency of 125 KHz. You wish to see what happens when you increase the frequency to 225 KHz and back again in steps of 10 KHz. Note Make sure that the waveform is a sinewave or squarewave before doing this example, otherwise the frequency set will be too invalid and an error will be caused.
  • Page 63: Example

    There are two ways to change a digit in the edit mode: Type in a new digit value by pressing the number on the numeric keypad. There is no need to press IEnter). Delete the digit before the cursor by pressing the key.
  • Page 64: Frequency

    Press IFreq) (frequency) to display the frequency of the Frequency output signal. To change this value either enter a new value with the numeric keypad, and press the relevent unit key; LMHz), ml, or @. edit the existing value using the edit function. Resolution of the frequency entry is 1 mHz for frequencies upto 999,999.999 Hz, independent of the waveform selected, and from 1 MHz upwards the...
  • Page 65: Example

    Example The following example shows how to change the frequency to 10.7854 KHz:- Result The Parameter display is shown. “FREQ” is high-lighted. “10.7854” is displayed. The frequency is then accepted and “10.7854 KHz” is displayed. Amplitude Press IAmpl (amplitude) to display the amplitude of the output signal.
  • Page 66: Amplitude Calibration

    Amplitude Calibration An amplitude calibration of the output signal can be performed on the HP 3324A. To do this, press the (Ampl) key (m a). The main signal output is switched off while the calibration is in progress.
  • Page 67: High-Voltage Option

    High-Voltage Option When the high-voltage output is used (option 002 is installed), a maximum output of 40 V peak-to-peak is available into a high impedance. The load resistance must be more than 50052 or distortion will result, particularly at higher frequencies. To assure square wave overshoot of <...
  • Page 68: Example

    Example The following example shows how to change the amplitude to 2.309 Vrms (6.531 V(p-p)) Keystrokes Result The Parameter display is shown. “AMPL” is highlighted. “2.309” is displayed. The amplitude is accepted and “2.309 V RMS” is displayed. DC Offset The m (DC offset) key enables you to display the DC offset of the output signal.
  • Page 69: Offset Only, No Ac Function

    Offset Only, No AC When the DC function is activated in the waveform menu (see Chapter 6), then no AC function is activated. Function The DC voltage output may then be programmed from 0 AC with DC Offset When DC offset is added to any AC function, there are minimum and maximum offset limits which must be observed.
  • Page 70 Table 5-4. Maximum DC Offset with any AC Function Attenuation Entry DC Offset DC Of&et Factor peak-to-peak) Entry (+ or -) 1.000 mV w i t h 4.500 A = 1000 0.001 mV 3.333 mV with 3.333 mV 3.334 mV with 14.99 mV A = 300...
  • Page 71: High-Voltage Option

    When the high-voltage output is used (option 002 is Option installed), the minimum and maximum permissible DC offset voltages may be determined by multiplying the amplitude and offset values in Table 5-4 by four. The equation given on the previous page must be changed to: Maximum DC offset = (20/A) - (Amptd/‘L) where A = Attenuation factor (from Table 5-4) and Amptd = Amplit,ude...
  • Page 72: Phase

    “EXT-REF” is shown in the display if an external reference frequency is used. Changing the phase of the HP 3324A will then cause the phase between the two instruments to change accordingly. Note In this case the phases of the two instruments are not calibrated, just locked.
  • Page 73: Example

    To change the phase, either enter a new value with the numeric keypad, and press the unit key (den. To enter a negative phase, press modify the existing value using the modify function. edit the existing value using the edit function. Note For squarewave frequencies below 25 KHz, phase changes greater than 25”...
  • Page 74: Automatic Phase Calibration

    Calibration respect to the other. To be able to do this it is necessary to have one of the HP 3324As installed with option 003 (slave), and the other with option 004 (master). If one of the instruments has option 001, High-Stability Frequency...
  • Page 75: Changing The Output Signal Cable Delay

    111’ 332412’s containing options 003 and 004. If the two cables a.re the same then in both HP 3324A’s the “Sig.d/ps:” zero. To account for a difference in the cables enter the delay in one instrument with respect to the other, that is enter the delay in the one with the longer cable.
  • Page 76 DUT with an oscilloscope. replace Keystrokes Result Press this on both HP 3324A’s, so that the utilities menu is accessed. Scroll through the menu on both HP 3324A’s until “set default” is shown. Scroll through the menu on both HP 3324A’s until “Sig.d/ps: Press this only on the slave instrument, so that an automatic phase calibration is performed.
  • Page 77: Changing The Calibration Cable Delay

    Changing the Calibration Cable Delay (003 or When an automatic phase calibration option 004) is installed “Cal.d/ps: n”, where n is a value from 0 to 59999, will be displayed in the UTILS menu. This can be used to account for the delay caused by changing the standard cable (supplied) for connecting the Phase Cal Out of the master (option 004) to the Phase Cal In of the slave (option 003).
  • Page 79: Waveforms Introduction

    Waveforms Introduction This chapter explains how to use the (E) key, which is one of the Menu keys. Press (jj to display the available waveforms, as shown in Figure 6-l. The meaning of each of the symbols can be seen in Figure 6-2. The waveforms always shown are the sinewave, squarewave, triangle, positive slope ramp, negative slope ramp, auxiliary TTL...
  • Page 80 T T L O u t p u t flux Figure 6-2. Waveforms Available Once a waveform has been activated, the HP 3324A jumps back to the point (menu) previous to entering the waveform menu, except for the following cases: the “AUX”...
  • Page 81: Example

    The auxiliary TTL clock waveform, chosen by “RUX”, is an output that is compatible to TTL levels. Into a into an open impedance, the high voltage-level is V. When “AUX” is selected, the main signal output is switched off. A TTL clock signal is output from the sync connectors for frequencies from 1 mHz to 60 MHz.
  • Page 83: Sweep

    Sweep Introduction Pressing the (G) key, one of the Menus keys, displays the Sweep menu, an example of which is shown in Figure 7-1. Mode I N T ERVRL CON’ M R R K E R STOP-FREQ SWEEP-TIME U s e d t o ed\t t h e P a r a m e t e r s t h e I n t e r v a l M a r k e r...
  • Page 84 This is also the case at power-on, if the HP 3324A is in the HP 3325A compatible mode. If not, the instrument is powered up to the settings that were at power-down. 7-2 Sweep...
  • Page 85 Table 7-1. Sweep Default State Parameter Mode Marker/Interval 10 (0 t,o 9) No. of stores Retrace Time Multi-Interval Sweep: 1 (liilear, all int,ervnls) Sweep Mode 1 M!lz (all intervals) Start Frequency Stop Frequency 5 Mlllz (all intervals) Marker Frequency 1 xc. (all inhvals) Sweep Time Activated Multi-Marker Sweep:...
  • Page 86: Multi-Interval Sweep

    Note Changing the number of intervals will cause the HP 3324A to reset all sweep intervals to their default state, and all stores to be cleared, so don’t program any intervals until you are sure you have enough available.
  • Page 87 Each interval is programmed with its own start frequency, stop frequency, sweep time, marker frequency, and whether its swept either linearly or logarithmically. means A linear interval may be either up or down, this that the stop frequency may be higher or lower than the start frequency.
  • Page 88 The marker output is a TTL-compatible output. It is high at the start of each interval, and when the marker frequency is reached it goes low, where it remains until the end of the interval. 50 individually-programmmed intervals may be used in a customized sweep-sequence of upto 100 intervals.
  • Page 89 The sweep sequence (which can consist of one interval) can be swept either singularly or continuously. A single sweep means that the HP 3324A sweeps from the start frequency of the first interval in the programmed sequence to the stop frequency of the last interval in the sequence, and then remains there.
  • Page 90: Setting Up A Multi-Interval Sequence

    This section will show how to set up a multi-interval Setting up a sweep sequence. Press lsweepl to show the Sweep menu. Multi-Interval To choose the multi-interval mode, put the cursor on Sequence “INTERVAL” and press (Select). A display similar to that in Figure 7-4 will be shown.
  • Page 91 When “INTERVAL” has been selected the current sweep sequence is shown. In the above example that is 1~3+6~2+9~4~9. This means that the sequence 1 3 4 5 6 2 3 4 5 6 7 8 9 4 9 is swept. m means “and then”, while (3) means “through”.
  • Page 92 To do this press the a key. You will then move onto some point in the list of start frequency, stop frequency, and so on. The position depends on where the cursor was when you last left and is shown by the parameter blinking.
  • Page 93: Example

    Example This example shows how to enter the sweep menu, and set up a continuous multi-interval sweep containing three sweep intervals. The intervals will be swept in the order 2, 3, 1, and the retrace time will be 5 s. The first thing to do is to enter the Sweep menu, choose the multi-interval mode, and set up the first interval, which will have a start frequency of 1 KHz, a stop...
  • Page 94 The following figure shows what the display will look like now : I N T E R V R L S I N G L E C O N T M A R K E R - F R E Q SWEEP-TIM Figure 7-5.
  • Page 95 The following figure shows what the display will look like now: STOP-FREQ MRRKER-FREO S W E E P - T I M E Figure 7-6. Example of 2nd Interval The third interval must now be programmed. It will have a start frequency of 5 KHz, a stop frequency of 1 KHz, a marker frequency of 3 KHz, a sweep time of 2 s and will be swept linearly.
  • Page 96 The following figure shows what the display will look like now: S I N G L E C O N T MARKER-FREQ SWEEP-TIM Figure 7-7. Example of 3rd Interval Now the sequence must be set, the continuous mode chosen with a retrace time of 5 s, and the sweep started. Keystrokes Enter The current retrace time is shown, and a continuous mode...
  • Page 97: Multi-Marker Sweep

    Multi-Marker The multi-marker sweep consists of only one linear interval, but upto 9 markers. The frequency for each Sweep marker is programmed separately. You can then activate the markers that you require. The marker pulsewidth is approximately 1 ms. For example consider having programmed the following interval.
  • Page 98 Figure 7-9. An Example Marker Sequence There is a minimal time difference of 1.5 ms between start- and marker-, marker- and marker-, and marker- and stop-frequencies. If a marker is too close to the start- or stop-frequency, it is not generated. If two markers are too close the second one (in time) is not generated.
  • Page 99: Setting Up A Multi-Marker Interval

    A continuous sweep means that the HP 3324A sweeps from the start frequency of the interval to the stop frequency, and then remains there for a programmed time (retrace time), before starting again at the start frequency.
  • Page 100 In the sweep display, two values in the form “n of 9” are shown (in the above example “n” = “1”). “n” means the number of the marker whose frequency you will see if you put the cursor onto the marker frequency. The “9” is the number of markers that can be programmed and activated in the multi-marker interval.
  • Page 101 Having chosen the marker number you wish to program, you can move onto changing the frequency. To do this press (5J. You will then move onto the marker frequency. You can then change the frequency by either: Typing in a new value with the numeric keypad, and then pressing the appropriate unit key.
  • Page 102: Example

    Example This example shows how to enter the sweep menu, and set up a single multi-marker sweep containing three markers. The start frequency will be 1 KHz, the stop frequency 10 KHz, and the sweep time 10 s. The markers will have the frequencies 3 KHz, 5 KHz and 7 KHz.
  • Page 103 The following display will be shown: SINGLE C O N T NRRKER-FREP SWEEP-TIME Figure 7-11. Marker Interval Example A single sweep has to be chosen now and the sweep started. Result Keystrokes T h e s w e e p s t a r t e d .
  • Page 104: The Sweep Keys

    This section explains the use of the sweep keys placed to The Sweep Keys the right of the front panel. Figure 7-12. The Sweep Keys Once a sweep has been defined it can then be started. This is done using the (ResetStartJ key. The (Start) key performs three functions for the sweep operations: (Reset) cancels the sweep.
  • Page 105: Example

    If a sweep has been stopped, reset or started can be seen on the right of the display, where an arrow is shown next to either “Reset”, “Run” or “Stop”. In the multi-interval linear sweep mode, the (afx2) bandwidth, respectively. If one of the absolute sweep frequency limits (0 and ‘21 MHz for a sinewave) is exceeded, the new interval is clipped.
  • Page 107: Utilities

    The instrument’s parameters are set to their default values if “set default” is displayed and m is ‘pressed. Parameters The default parameters are shown in Table 8-1. Note Setting the HP 3324A to default will cause the instrument to clear all its stores. Utilities 8-1...
  • Page 108 Table 8-1. Parameter Default Values Default Parameter Function Sine Frequency 1000 Hz Amplitude Phase 0 degrees DC Offset Selected Output 1 (front) output Marker/Interval Mode 1 (multi-interval) No. of stores 10 (0 to 9) Retrace Time 0 sec. (auto-retrace) Multi-Interval Sweep: Sweep Mode 1 (linear, all intervals) Start Frequency...
  • Page 109: Choosing The Number Of Stores/Intervals

    40 - 50 Note Changing either the number of stores or intervals will cause the HP 3324A to reset all sweep intervals to their default state, and all stores to be cleared. Recommendation: Use this command sparingly and only when absolutely necessary.
  • Page 110: Changing The Hp-Ib Address

    HP-IB address. The default HP-111 address is 17. Choosing HP The HP 3324A can be set so that at power-on, it powers up just like the HP 3325A. This means that in the 3325A compatible mode the main output signal is enabled and...
  • Page 111: Performing A Self-Test

    The “all” test is also an interactive test. If an error is shown return the instrument to an HP Service Center for repair. To perform one of the self-tests, display “selft: std./all”...
  • Page 112: Changing The Output Signal Cable Delay

    IIP 3324A’s containing options 003 and 004. If the two cables are the same then in both HP 3324A’s the “Sig.d/ps:”...
  • Page 113: Checking The Software Revision

    Checking the In the utilities menu, one of the parts of the menu shows the software revision incorporated in the HP 3324A. Software Revision Scroll the menu until “ROM rev. : x.x” is displayed. This is the software revision. Switching the If the high-voltage option 002 is installed, it can be either enabled or disabled as the main output signal.
  • Page 115: Programming The Hp 3324A

    A complete syntax list of the HP 3324A programming commands can be found in Appendix E. The HP 3324A The HP 3324A interfaces to the HP-IB as defined by the IEEE Standard 488.1. The interface functional subset HP-IB Capabilities which the HP 3324A implements is specified in Table 9-l.
  • Page 116: The Interpreter

    HP 3325A. Note Execution times for the HP 3324A HP-IB commands are slower than those for the HP 3325A. To make sure that a command has been processed, check the “Busy” bit in the status byte. Chapter 11 and Appendix E show which HP 3324A commands are compatible to the HP 3325A and which are not.
  • Page 117 Frequency 10 MHz Stop Frequency 5 MHz (all markers) Marker Frequency 1 sec. Sweep Time I (only the first marker) Activated Markers (unbuffered) Data Transfer Mode SRQ Mask FR (see section “Omitting Default Mnemonic Mnemonic”) Programming the HP 3324A 9-3...
  • Page 118: Hp-Ib Display Indicators

    HP-IB Display The following list explains the indicators that can be seen in the display when the HP 3324A is being Indicators controlled remotely. “REMOTE” “REMOTE” is displayed when the HP 3324A is operating under remote control. While in remote (and local lockout is not in effect), only the @ key is recognized.
  • Page 119: Accepted Characters

    Table 9-3. Accepted Characters Description Character White spaces HT, LF, CR, space Block terminator Plus sign Minus sign / Range separator Item separator Decimal point Digits / Integer List terminator Letters Programming the HP 3324A 9-5...
  • Page 120: Hp-Ib Messages

    HP-IB Messages Table 9-4 shows which HP-IB messages are implemented/not implemented in the interpreter. Table 9-4. HP-IB Messages Treatment Command Data byte Implemented as usual Device clear Sets 3324 into reset state, see section “The Interpreter” End or identify End is treated as white space...
  • Page 121: Command Groups And Data Transfer Modes

    Command Groups The HP 33‘24A works in two separate data transfer modes. In mode 1, the unbuffered mode, commands and Data Transfer are executed as soon as they are received. In mode Modes 2, the buffered mode, the execution control depends on the command group to which a command belongs.
  • Page 122 Treatment Group Executed immediately, HP 3324A is put into mode 2. Group 0 Group 1 Command is remembered but not executed. Group 1 Command is remembered but not executed. Group 0 Commands “FUl” and “FR12.3MH” are executed, along with command “PH5DF.
  • Page 123 In data transfer mode 1, the only command that will not be executed is the string The following events will cause remembered, but not executed commands to be lost: power off remote commands. Programming the HP 3324A 9-9...
  • Page 124: Omitting The Mnemonic

    1, and 3 MHz. This could be done in the following way: Send program string Do measurement Send program string “2MH” Do measurement Send program string “3MH” Do measurement 9-10 Programming the HP 3324A...
  • Page 125: Synchronizing After A Syntax Error

    To change the address, type in the required address with the keypad, and then press IEnter). The HP-IB refuses to accept a new HP-IB address if it is either in the remote state, or if it is addressed either as a listener or as a talker.
  • Page 127: Status Reporting

    Status Reporting Introduction Another important feature of the III’ 3324A is that you can program it to interrupt the controller when certain status/error conditions are met. An 8-bit register, the Status Byte, contains the instrument’s status. In order for an interrupt to occur, it must be programmed via the Service Request Enable Mask.
  • Page 128 7 Bit 5 Bit 4 Bit. 3 Bnt 2 Bit L L-- I =Prog P am s:r',b"p I =Sweep Stopped S t a r t e d I-System Always 0 I ” P r o g r e s s Reqest Busy Figure 10-l.
  • Page 129 Sweep. Set when a sweep is in progress, clear when a sweep is in progress. Cannot be configured to cause an SRQ. Require Service. Set when the HP 3324A requires service (sent SRQ). 1 2 8 Busy. Set while a command is being executed, clear when instrument is not busy.
  • Page 130 The program string error bit is set when a command string is received that the interpreter does not understand, or when a command is not executable for When the “IER” query some reason. is sent, an error number will be returned, with the following meanings: Error Error Description...
  • Page 131 The system failure bit is set when a system failure is encountered. The specific error may be found by sending the “ISE” query, and evaluating the returned error number: System Failure Description System Failure Number Amplitude calibration failed Phase calibration failed External reference unlocked Main oscillator unlocked Self-test failed (report actual number to...
  • Page 132: Forcing An Srq

    Forcing an SRQ An SRQ is forced when one of the above bits in the Status Byte Register goes from “0-> 1” AND the corresponding Service Request Enable Mask bit has been set. Bits O-3 are never set if they are not enabled via the SRQ-mask command.
  • Page 133: Service Request Enable

    Service Request The SRQ mask can be set to mask bits O-3 of the Status Register. To set the mask first determine which Enable conditions you want to interrupt the controller, for example sweep started, sweep stopped and so on. Determine the decimal number (1-16) corresponding to those conditions.
  • Page 134: Example 1

    Example 1 The following example shows how to set the SRQ mask, so that if a “0 + 1” transition of the Sweep Started or System Failure bit occurs, an interrupt is sent. S y s t e m Sweep A l w a y s S y s t e m Sweep Sweep...
  • Page 135: Example 2

    Example 2 order to see when a sweep has started. Result Program Clears all devices. HP-IB SEND 7; CMD 20 Sets SRQ mask for bits 2 and 3 of the status OUTPUT 717; "MSL" byte. Serial polls the status byte.
  • Page 137: Hp-Ib Commands And Programming Examples

    Program end. The first thing to check if the amplitude is not shown, is that the HP-IB address set on the HP 3324A is the same as that you are using (in this case 17). HP-IB Commands and Programming Examples...
  • Page 138 Table 11-l. HP 3324A Command Summary S R O - 9 Recall set SNR Integer White C M O - o f f node 1 - on Function F U O - D C 1 - Sine 2 - Square...
  • Page 139 Table 1 l-l. HP 3324A Command Summary (continued) 0 phase Select R F 0 - f r o n t o u t p u t 1 - rear o u t p u t O O F O - o f f...
  • Page 140 Table 11-l. HP 3324A Command Summary (continued) Data 3325 Multi-Interva start freq Number See FR Index, number stop freq Number Index, number marker freq Number Index, number sweep time Number SE - sec. Index, number sweep mode 1 - lin...
  • Page 141: Notes

    N o t e s (1) The meaning of the command has been changed. See the section “Command Groups and Data Transfer Modes” in Chapter 9. The data transfer command, MD, allows you to select between a buffered and an unbuffered transfer mode. In the unbuffered mode, the command is executed immediately.
  • Page 142 33.33 mV with 33.33 mV 149.9 mV 33.34 mV with 0.010 mV 99.99 mV with 116.6 mV with 450.0 mV 100.0 mV 0.100 mV 333.3 mV with 333.3 mV 333.4 mV with 1.499 V 11-6 HP-IB Commands and Programming Examples...
  • Page 143 9. The list must define the subset of enabled markers (not the sequence). All list elements must be in the range to 9. The default list is “1” (only one marker, the first one). HP-IB Commands and Programming Examples 11-7...
  • Page 144 In all other cases, the retrace time is evaluated according to (13). The output frequency sweeps down/up from the stop frequency of the last interval in the sweep sequence to the start frequency of the first interval. 11-8 HP-IB Commands and Programming Examples...
  • Page 145 (12) If the total sweep time (for multi-intervals, this is the sum over all selected intervals) is greater than 99.99 seconds, then the X-drive is switched off. This means that the X-drive output is held low all of the time. The Z-blank output, however, works as usual.
  • Page 146: Programming Examples

    All of the programs assume the following: an HP 9000, Series 200 or 300 Computer as controller 2. that BASIC is the programming language 3. that the HP 3324A is preset to HP-IB address 17 Function Generator The following program shows an example of how to program the HP 3324A as a function generator.
  • Page 147: Multi-Interval Sweep Generator

    Multi-Interval Sweep The following program shows an example of how to program the HP 3324A as a multi-interval sweep Generator generator. The program assumes that the amplitude, waveform, data transfer mode, DC offset, phase, output and no. of intervals are the same as set in the previous example.
  • Page 148: Program

    Sets retrace time of 1 sec., sets multi-interval OUTPUT 717; "RTTISE MHSl mode and sets the interval sequence 1 3 5 2 4 5 1 2 3 4 5 . Sets continuous sweep mode. OUTPUT 717; "SC *" Progam end. 11-12 HP-16 Commands and Programming Examples...
  • Page 149: Multi-Marker Sweep Generator

    Multi-Marker Sweep The following program shows an example of how to program the HP 3324A as a multi-marker sweep Generator generator. The program assumes that the previous two example programs have been run directly beforehand. The program sets the following information.
  • Page 151: Performance Tests

    Introduction The procedures in this chapter test the performance of the instrument. The complete specifications to which the HP 3324A is tested are given in Appendix A. All tests can be performed without access to the interior of the instrument.
  • Page 152: Test Failure

    If the HP 3324A fails any performance test, return the Test Failure instrument to the nearest Hewlett-Packard Sales/Service Office for repair. Specifications are the performance characteristics of the Instruments instrument which are certified. These specifications, Specification listed in Appendix A, are the performance standards or limits against which the HP 3324A can be tested.
  • Page 153 Table 12-1. Recommended Test Equipment Recommended Model Instrument Critical Specifications HP 1725A Vertical A n a l o g Oscilloscope Bandwidth: DC to 275 MHz Deflection: 0.01 to 5 V/div Horizontal Sweep: 10 ns to 0.5 s/div Delayed Sweep HP 54111D...
  • Page 154 Ranges: 0.1 to 1 V HP 400FL/3400A AC Voltmeter Frequency Range: 20 Hz - 1 MHz Input Impedance: 11 MR Meter: Log scale Frequency: 20 KHz HP 3324A/3325B Sincwave Signal Source Amplitude: 10 V(p-p) into 500 HP 6214A/6214B DC Power Supply Amps: 10 mA...
  • Page 155: Harmonic Distortion

    Verify that all harmonics are at least 25 dB below the fundamental. 4. Set the HP 3324A to 15 MHz and verify that all harmonics are at least 25 dB below the fundamental. Performance Tests 12-5...
  • Page 156: High-Voltage Output (Option 002)

    1.99 MHz 200 KHz - 4 0 dB 6. Set the HP 3324A frequency to 50 KHz and the amplitude to 9.99 mV(p-p). 7. Set the spectrum analyzer controls to display the fundamental and at least three harmonics. (It may be necessary to decrease the video bandwidth to separate the harmonics from the noise floor.) Verify that all...
  • Page 157 4 7 5 O h m s Figure 12-1. Harmonic Distortion Verification Test Set-Up (High-Voltage Output) 2. Select the high-voltage output on the HP 3324A. Set the amplitude to 40 V(p-p) and the frequency to 100 3. Set the spectrum analyzer controls to display the fundamental and at least three harmonics.
  • Page 158: Spurious Signal

    Equipment Required Spectrum Analyzer Mixer Spurious 1. Connect the HP 3324A signal output to the spectrum analyzer 500 (R,F) input and the III’ 3324A EX’I Procedure REF input to the analyzer 10 MHz reference output, as shown in Figure 12-2.
  • Page 159 25.998 MHz to display the 3:2 mixer spur. Verify that this spur is at least 55 dB below the fundamental. 7. In a similar manner, change the HP 3324A frequency and the spectrum analyzer center frequency to the following frequencies. For each setting, verify that all spurious signals are 55 dB below the fundamental.
  • Page 160: Close-In Spurious (Fractional N Spurs) Procedure

    Close-in Spurious This procedure continues on from the previous one. (Fractional N Spurs) 1. Set the HP 3324A to 5.001 MHz and the amplitude to Procedure 448.3 mV(p-p). 2. Set the spectrum analyzer controls as follows: 5.001 MHz Center Frequency...
  • Page 161: Integrated Phase Noise

    1 MHz Low Pass Filter 1. Connect the equipment as shown in Figure 12-3, Procedure connecting the 15 KHz noise equivalent filter output to the AC voltmeter. Phase lock the HP 3324A and the signal generator together. HP 3324A Voltmeter...
  • Page 162 5. Change the HP 3324A frequency to 19.9 MHz. 6. Connect the 15 KHz filter output to the digital voltmeter. 7. Press the HP 3324A (Phase) key. Using the modify keys, adjust the output phase for minimum reading on the digital voltmeter.
  • Page 163: Ramp Retrace Time

    Ramp Retrace This procedure tests the HP 3324A retrace time of the positive and negative slope ramps. Time Specifications Equipment Required Digitizing Oscilloscope Procedure 1. Connect the HP 3324A signal output to the oscilloscope vertical input. (If your oscilloscope does not have a 5052 input, use a 5052 feedthru termination at the input.)
  • Page 164: Frequency Accuracy

    Equipment Required Electronic counter (calibrated within three months or with an accurate 10 MHz external reference input) 1. Connect the HP 3324A signal output to the electronic Procedure counter channel A input with a 500 fecdthru termination. Allow the HP 3324A t,o warm up for 20 minutes and the counter’s frequency reference to...
  • Page 165: Phase Increment Accuracy

    Set the counter to average 1000 periods. The electronic counter should indicate 6. Change the HP 3324A function to a positive slope ramp. The electronic counter should indicate This procedure compares the phase increment accuracy Phase Increment to the specification.
  • Page 166 Slope A and 1~ Positive Sample Rate Maximum 5. Press the III’ 3324A m key to display phase. Using the modify keys, adjust the pha.se until the counter reads approximately 200 ns. Press the blue (Shift key, and then the key.
  • Page 167 8 and 6, and record in the Time Difference column. The difference should be f0.5”. 10. Set the HP 3324A phase to -10”. 11. Reset the counter. Record the counter reading in the space for 10’ Increment Time Interval.
  • Page 168: Amplitude Accuracy

    Note For each new setting of the amplitude, press @&i-G” to perform a.n amplitude calibration. 1. Sinewave Test. Connect the HP 3324A signal output Amplitude Accuracy through a 50R feedthru termination to the AC at Frequencies upto digital voltmeter input.
  • Page 169 5. Change the HP 3324A amplitude to 1.061 Vrms (3 KHz and 100 KHz as above. Verify that all three voltmeter readings are between 1.037 and 1.085 Vrms (ho.2 dB). 6. Change the HP 3324A amplitude to 0.3536 Vrms (1 V(p-p)) and set the DC offset to 1 mV. Set the HP 3324A frequency to 100 Hz, 1 KHz and 100 KHz, and read the AC voltage.
  • Page 170 Add the two readings to obtain volts peak-to-peak. Verify that the sum is between 9.85 and 10.15 V. 14. Change the HP 3324A function to a triangle. 15. Adjust the oscilloscope delay to read the positive peak voltage on the high-speed digital voltmeter.
  • Page 171 Verify that the sum of the absolute values is between 9.85 and 10.15 V. 22. Change the HP 3324A function to a triangle and the frequency to 2 KHz. Set the oscilloscope main sweep to 50 PsJdiv. Adjust the delay and position. Set the positive and negative trigger to read the peaks.
  • Page 172 23. Change the HP 3324A function to a positive ramp and the frequency to 500 Hz. Set the oscilloscope main sweep to 0.2 ms/div (adjust the sweep vernier to return the peaks to the center screen). Verify that the voltage is between 9.85 and 10.15 V.
  • Page 173 29. Change the HP 3324A amplitude to 3 V(p-p), and press [mj. Set the HP 3324A frequency to 99.9 Hz and the function to squarewave. 30. Repeat tests 9 through 28. Test limits are as follows: Function 99.9 Hz Square 3.045 v...
  • Page 174 31. Change the HP 3324A amplitude to 1 V(p-p), press frequency to 99.9 Hz and the function to square. Set the oscilloscope’s vertical sensitivity to 0.05 V/div for all 1 V(p-p) tests. 32. Repeat tests 9 through 28. Test limits are as follows:...
  • Page 175: High-Voltage Output (Option 002)

    AC voltmeter with a 6-foot cable. Connect a line. Select the high-voltage output on the HP 3324A. “HIGH-VOLT” is shown in the display. Set the HP 3324A function to a sinewave, the 14.14 frequency to 2 KHz, and the amplitude to Vrms reading should be between 13.86 and 14.42 Vrms.
  • Page 176 Connect the tap to the input of the high-speed voltmeter, as shown in Figure 12-6. 6. Set the HP 3324A frequency to 2 KHz and the amplitude to 40 V(p-p). Set the voltmeter to the 1 V range and external trigger. Set the oscilloscope as...
  • Page 177: Amplitude Flatness (Frequencies Above 100 Khz)

    Frequency DC Offset 10 dBm Amplitude 2. Connect the HP 3324A signal output via a 500 to measure AC voltage. 3. Press [Ampl followed by C-Cal). Wait 2 seconds and record the voltmeter reading as Vnrms. 4. Connect the equipment as shown in Figure 12-7 (the...
  • Page 178 Calculate the amplitude accuracy using the following equation: 8. Verify that all results are within 9.4 dBm and 10.6 9. Change the HP 3324A frequency to 1 KHz and the amplitude to 18 dBm. 10. Repeat steps 2 to 7 and verify that all results are within 17.6 and 18.4 dBm (accuracy = 60.4 dB).
  • Page 179: High-Voltage Output (Option 002)

    Sensitivity 10 V/div Time/div 1 ms 3. Set the HP 3324A to 40 V(p-p) sinewave and 1 KHz. Adjust the oscilloscope intensity and focus for a sharp trace. 4. Use the modify keys to increase the HP 3324A frequency from 1 KHz to 1.001 MHz in 200 KHz steps.
  • Page 180: Dc Offset Accuracy (Dc Only)

    Connect the HP 3324A signal output directly to the the DC digital voltmeter input. 3. Set the HP 3324A DC offset to 5 V, then press 4. The voltmeter reading should be between i-4.950 and 5. Change the HP 3324A DC offset to -5 V. The voltmeter reading should be between -4.950 and...
  • Page 181: Attenuator Test

    HP 3324A output directly to the voltmeter input. 2. Select the high-voltage output on the HP 3324A. “HIGH-VOLT” will be shown in the display. 3. Set the HP 3324A DC offset to 20 V. The voltmeter reading should be +19.775 to 20.225 V. Performance Tests 12-31...
  • Page 182: Dc Offset Accuracy With Ac Functions

    DC Offset This procedure tests the HP 3324A DC offset accuracy when an AC function output is present. Accuracy with AC Functions Specifications DC + AC, upto 1 MHz: 1.2% For ramps upto 10 KHz: 2.4% DC + AC, from 1 MHz to 20 MHz: 3%...
  • Page 183 5. .Change the HP 3324A frequency to 999.9 KHz. The voltmeter reading should be -4.440 to -4.560 Vdc. 6. Change the HP 3324A DC offset to +4.5 V. The voltmeter reading should be +4.440 to +4.560 Vdc. 7. Set the HP 3324A function to square. The voltmeter reading should be -f-4.440 to i-4.560 Vdc.
  • Page 184: Triangle Linearity

    Triangle Linearity This procedure tests the linearity of the HP 3324A triangle wave output. As the triangle and ramp outputs are generated by the same circuits, this procedure effectively tests the ramp linearity also. Specifications Equipment Required High-speed DC Digital Voltmeter...
  • Page 185 Number of readings Trigger External Note The HP 3437A triggers on the negative going edge of the HP 3324A sync squarewave. 4. Set the voltmeter delay to 0.00003 (seconds). Record the voltmeter reading on the Performance Test Positive Slope Measurement, (lO%)yl.
  • Page 186 6. Measure the voltage at each 10% segment point on the negative slope by setting the voltmeter delay to the following. Enter on the Performance Test Record in the appropriate spaces under Negative Slope Measurement. Percent of Slope Delay 0.00008 0.000085 0.00009 0.000095...
  • Page 187 11. First determine the value of al using the following equation: where Cx, Cy, Cxy, CxCy, ,Xx2, and (CX)~, are the previously calculated values entered on the Performance Test Record. where n=9 (the number of points to be calculated). 12. Determine the value of ac using the equation: 13.
  • Page 188: Ramp Period Variation

    The voltages measured and recorded in the within the calculated tolerances. Ramp Period This procedure tests the variation between alternate cycles of the HP 3324A positive and negative slope Variation ramps. Specifications < kl% of period, maximum...
  • Page 189 6. The reset portion should show more than one line, as in Figure 12-9. The lines should not be separated by more than five divisions on the display. 7. Change the HP 332412 function to positive slope ramp and set the oscilloscope trigger to negative to verify the positive ramp.
  • Page 190: Automatic Phase Calibration

    > 1 MHz - 10 MHz ~t1.2~ , tYP. Digitizing Oscilloscope Equipment Required Note Connect a 332442 signal to the oscilloscope and using 15109, an adder (splitter) HP check the oscilloscope interchannel accuracy and take the error into account. 12-40 Performance Tests...
  • Page 191: Procedure

    CRL IN 0 0 0 3 SLRVE BNC CRBLE 500HM S A M E L E N G T H Figure 12-10. Phase Calibration Test Set-Up 2. Set the two HP 3324As as follows: Util menu Set default Frequency 1 MHz...
  • Page 192 # Avgs Press (Edge). Check that & < 4.17 ns. Change the waveform of the two HP 3324As to square. Select the Delta V menu on the scope and with preset levels at 50-50% press Auto Level Set.
  • Page 193: Test Record 7

    Test Record 1 Hewlett-Packard Model 332411 Teat Performed By Synthesized Function/Sweep Generator Date Serial Number Comments CSO No. Pass 0 Fail 0 Harmonic Distortion Test Specification Fundamental F’requency -25 dB 20 MHz 15 ME&z 14.9 MHz -30 dB 2 MHz 1.99 MHz 200 KHz 10 KHz...
  • Page 194: Test Record 8

    Test Record 2 Hewlett-Packard Model Test Performed By 3324A Date Synthesized Function/Sweep Generator Comments Serial Number CSO No. Customer Spurious Signal Test Pass 0 Fail 0 Spurious Mixer 2.001 4.100 MHz 6.100 MHz 8.100 MHz 10.100 MHz 12.100 MHz 14.100 MHz 16.100 MHz...
  • Page 195: Test Record 8

    Test Record 3 Test Performed By Synthesized Function/Sweep Generator Date Comments Serial Number CSO No. Customer Integrated Phase Noise Test Pass 0 Fail 0 1st AC voltmeter reading 2nd AC voltmeter reading Specification Result -- 6 dB - - - - - - - - - - - - - - - - - - - - - - - - Pass 0 Fail 0 Specification Positive Slope Ramp...
  • Page 196: Test Record 9

    Test Record 4 Hewlett-Packard Model 3324A Test Performed By Synthesized Function/Sweep Generator Date Serial Number Comments Customer CSO No. Frequency Accuracy Test Pass 0 Fail 0 Sine, 20 MHz Square, 10 MHz - - - - - - - - - - - - - - - - - - - - - - - - -...
  • Page 197 Test Record 5 Hewlett-Packard Model 3324A Test Performed Hy Synthesized Function/Sweep Generator Date Serial Number Comments Customer CSO No. Amplitude Accuracy Test Pass 0 Fail 0 Measured Amplitude: 3.536 Vrms Sine, 100 Hz: 3.455 v 3.617 V Sine, 1 KHz: 3.455 v...
  • Page 198 (2 of 4) Test Record 5 Amplitude Accuracy Test (continued) Function Test Measured Amplitude: 10 V(p-p) Square, 99.9 Hz 10.15 v 9.95 v Triangle, 99.9 Hz 10.15 v 9.85 v Pos. Ramp, 99.9 Hz 10.15 v 9.85 v Neg. Ramp, 99.9 Hz 10.15 v 9.85 v 10.15 v...
  • Page 199 Test Record 5 Amplitude Accuracy Test (continued) Function Test Measured Amplitude: 1 V(p-p) DC: 1 mV Square, 99.9 Hz 1.022 v 0.978 V 1.027 V Triangle, 99.9 Hz 0.973 v 1.027 V Pos. Ramp, 99.9 Hz 0.973 v 1.027 V Neg.
  • Page 200 Test Record 5 Accuracy Amplitude Test (continued) Min. Amplitude Flatness Test 10.2 dBm 9.8 dBm Sine, 10 dBm, 1 KHz 10.6 dBm 9.4 dBm 101 KHz 10.6 dBm 9.4 dBm 500 KHz 10.6 dBm 9.4 dBm 1 MHz 10.6 dBm 5 MHz 9.4 dBm 10.6 dBm...
  • Page 201 Test Record 6 Test Performed By Hewlett-Packard Model 3324A Function/Sweep Generator Date Comments Serial Number Customer CSO No. Pass 0 Fail 0 DC Offset Accuracy Test (DC Only) Minimum Maximum Measured -5 v -4.950 v - - - - - - - - - - - - - - - - -5.050 v 1.499 v...
  • Page 202 Test Record 7 Test Performed By Hewlett-Packard Model 3324A Synthesized Function/Sweep Generator Date Serial Number Comments Customer CSO No. Offset Accuracy Test with AC Functions Pass 0 Fail 0 Measured Maximum Minimum Sine, 21 MHz 4.5 v -4.5 v -4.350 v -4.650...
  • Page 203 Test Record 8 Test Performed By Hewlett-Packard Model 3324A Date Synthesized Function/Sweep Generator Comments Serial Number CSO No. Customer Pass 0 Fail 0 Triangle Linearity Test Minimum Maximum Calculated x Values Positive Slope x times y Tolerance Tolerance Best Fit...
  • Page 204 Test Record 8 Pass 0 Fail 0 Triangle Linearity Test (continued) Minimum Maximum Calculated Negative Slope Tolerance Tolerance Best Fit Measurement Straight Line (10%) y1----- - -- - - - - - - (20%) y2----- x3 = 3 (30%) y3----- (40%) y4----- x5 -= 5 (50%) y5-----...
  • Page 205 Test Record 9 3324A Test Performed 13~ Synthesized Function/Sweep Generator Date Comments CSO No. Automatic Phase Calibration Test Pass 0 Fail 0 Formula: At < 4.17 Sine, 1 MHz At < 2.78 ns 1 MI12 Performance Tests 12-55...
  • Page 207: Waveforms

    0°C to 55°C ambient temperature. All specifications describe the warranted performance, except for the Main Signal Output, Squarewave Characteristics (also by Option 002), Auxiliary Outputs, Auxiliary Inputs, HP-IB Control and General specifications, which describe the typical performance. Waveforms Sine, square, triangle, negative and positive ramps, DC and TTL clock.
  • Page 208: Main Signal Output (Supplementary)

    Accuracy: calibration with standard frequency reference. Stability: option 001, high-stability frequency reference). Warm-up Time: 20 minutes to within specified accuracy. Impedance: Main Signal output Return Loss: (Supplementary) > 3 V, 5 MHz to 20 MHz. Floating: Output may be floated upto 42 V peak (AC + DC). Connector: BNC, switchable between front and rear panel.
  • Page 209 1 mV - 2.999 mV 100 mV - 299.9 mV 3 mV - 9.999 mV 300 mV - 999.9 mV 10 mV - 29.99 mV 1 v - 2.999 v 30 mV - 99.99 mV 3 v - 10.00 v Resolution: 4 digits (0.03% of full range).
  • Page 210: Sinewave Spectral Purity

    Squarewave: 0.001 Hz 100 KHz >lOO KHz 10 MHz Triangle: 0.001 Hz to 2 KHz >2 KHz to 10 KHz Ramps: 0.001 Hz to 500 Hz >500 Hz to 10 KHz With DC offset, increase all sinewave tolerances by 0.2 Phase Noise: -50 dBc for a 30 KHz band centered on a 20 MHz Purity...
  • Page 211: Squarewave Characteristics

    Harmonically related signals will be less than the following levels relative to the fundamental: Frequency Range Harmonic Level 0.1 Hz - 199 KHz -60 dBc 200 KHz - 1.99 MHz -40 dBc 2 MHz - 14.9 MHz -30 dBc 15 MHz - 20 MHz -25 dBc Rise/Fall Time:...
  • Page 212: Dc Offset

    DC Offset DC only (no AC signal): 0 to f5 V / 50Q DC + AC: Maximum DC offset f4.5 V on highest range; decreasing to f4.5 mV on lowest range. Resolution: digits Accuracy: DC only: f0.015 mV to f50 mV, depends on offset chosen, ~4~0.02 mV.
  • Page 213: Frequency Sweep

    Sequence Modes: Frequency Sweep Sweep Single, continuous. Sweep Function Modes: Multi-Interval: 50 different intervals can be sequenced and repeated in any order in a sequence which can contain Frequency-switching-time between intervals (to within 1 Hz): Multi-Marker: sweep (no Multi-Interval) is performed. Reaching a marker frequency generates a high-to-low transition at the marker output.
  • Page 214: Auxiliary Outputs (Supplementary)

    Auxiliary Outputs SYNC Output: 1 mHz to 21 MHz phase synchronous squarewave with (Supplementary) the same frequency as the main signal output, or 1 mHz to 60 MHz Auxiliary TTL Clock (main signal output switched off). Output impedance: 5051 Output levels: high level > 1.2 V, low level < 0.2 V Connector: BNC, front and rear panels.
  • Page 215: Auxiliary Inputs (Supplementary)

    Output amplitude: 0 dBm Connector: rear panel BNC. Reference Input Auxiliary Inputs For phase locking the HP 3324A to an external (Supplementary) frequency reference. Signal from 0 dBm to 20 dBm into 500 Reference signal must be a sub-harmonic of 10 MHz from 1 MHz to 10 MHz.
  • Page 216: Reference

    25OC, for an off time of less than 24 hours. 10 MHz Oven Output: 10 MHz squarewave for phase-locking additional instruments to the HP 3324A. Output impedance: 50R Output level: >4.5 dBm Connector: rear panel BNC.
  • Page 217: Automatic Phase Calibration Options

    HP 3324As without Options external measuring instruments. Option 003 has to be installed into one HP 3324A, option 004 has to be installed into the other instrument. Adjust factors for different cable delays between master/slave and can be entered in the UTILit,y menu.
  • Page 218 Phase Error The phase error is measured between the main signal outputs on the rear panel with either unequal or equal amplitudes and without DC offset. Phase is defined as the difference in rising edge to rising edge (measured with AC-coupled zero-crossing-points as reference points) for sine and squarewaves.
  • Page 219: General (Supplementary)

    Operating Environment: General Temperature: 0°C to 55”C, if not otherwise stated. (Supplementar Relative humidity: 95%, 0°C to 40°C Storage temperature: -40°C to +75”C Power: V, &lo%; 48 to 66 Hz,; 100 VA max. Standby 20 VA max. Weight: 11 kg net, 16.5 kg shipping. Dimensions: 132.6 mm high x 425.5 mm wide x 497.5 mm deep (5.25 ins.
  • Page 221: Introduction

    This appendix gives an explanation of the options and accessories available for the HP 3324A. Table B-l lists the options available for the HP 3324A. These options are available when the instrument is ordered by specifying the option number, or are available later by ordering the option part number (p/n).
  • Page 222: Reference

    001 requires 15 minutes of warm-up time to meet frequency specifications if power is disconnected for less then 24 hours. For more than 24 hours, the HP 3324A may require upto ‘72 hours warm-up time to meet the frequency specifications.
  • Page 223: Option 002 High-Voltage Output

    4 V(p-p). Turning the high-voltage off causes the output to be amplitude The HP 3324A specifications apply when the external load is 500R and the total capacitance is <500 pF. The same entry procedures and display features apply as for the standard configuration.
  • Page 224: Options 003 And 004 Automatic Phase Calibration

    To be able to do this it is necessary Phase Calibration to have one of the HP 3324As installed with option 003 (slave), and the other with option 004 (master). If one of the instruments has option 001, High-Stability Frequency...
  • Page 225 l CWL IN Figure B-l. Automatic Phase Calibration Setup Pressing the (Phase] key on the slave (the one with option 003 installed) will cause a phase calibration between the two instruments to be performed. “phase is sl~own shortly in the display of the slave, before displaying what was previously shown.
  • Page 227: Introduction

    Safety The Model HP 3324A is a Safety Class 1 instrument (instrument with an exposed metal chassis that is Considerations directly connected to earth via the power supply cable).
  • Page 228: Initial Inspection

    Power Requirements Caution Before applying AC line power to the HP 3324A, ensure that the voltage selector on the HP 3324A bottom panel is set for the proper line voltage and the correct line fuse is installed in the fuse holder. Procedures for changing the line voltage selector and fuse are contained in the following section “Line Voltage Selection”.
  • Page 229: Line Voltage Selection

    The line voltage selected for the HP 3324A is indicated by the switches on the bottom panel. Refer to Table C-l for the line voltage ranges and Table C-2 to set the line voltage and select the appropriate fuse.
  • Page 230 To change the line voltage and fuse: 1. Remove the power cord. 2. To check or replace the fuse, press the fuse holder in slightly, using a screw driver, and turn it to release the catch. Pull out the fuse holder. 3.
  • Page 231: Power Cable

    Power Cable In accordance with international safety standards, this instrument is equipped with a three-wire power cable. When connected to an appropriate AC power receptacle, this cable grounds the instrument cabinet. The type of power cable shipped with each instrument depends on the country of destination.
  • Page 232 Denmark Europe G r e a t Austral I a 8 1 2 0 - 1 8 120-2956 8 120- 2 0 - 1 3 5 1 8128-1378 S w i t z e r l a n d Figure C-2. Power Cables - Plug Identification The following work should be carried out by a qualified electrician - all local electrical codes being strictly observed.
  • Page 233: Hp-Ib Connector

    HP part number 1251-0293. HP-IB Logic Levels The HP 3324A HP-IB lines use standard TTL logic, the levels being as follows: True = Low = digital ground or 0 Vdc to 0.4 Vdc, False = High = open or 2.5 Vdc to 5 Vdc.
  • Page 234: Operating Environment

    S Y S T E M C O N T R O L L E R ) Figure C-3. HB-IB Connector The following summarizes HP 3324A operating Operating enviroment ranges. In order for the IIP 3324A to meet Environment specifications, the operating enviroment must be within these limits. Warning The HP 3324A is not designed for outdoor use.
  • Page 235: Instrument Cooling

    BNC cable. For example, use HP p/n 5180-2459 (1.22 m, RG58V Triax, 500). Instrument Cooling The HP 3324A is equipped with a cooling fan mounted inside the rear panel. The instrument should be mounted so that air can freely circulate through it.
  • Page 236: Storage And Shipment

    If the instrument is to be shipped to a Hewlett-Packard Return Shipments Sales/Service Office, attach a tag showing owner, return to HP address, model number and full serial number and the type of service required. The original shipping carton and packing material...
  • Page 237: Hp-Ib Overview

    Introduction This appendix gives an overview of the Hewlett-Packard Interface Bus (HP-IB). Description of the The HP-IB is a bus structure that links the HP 3324A to desktop computers, minicomputers, and other HP-IB HP-IB controlled instruments to form automated measurement systems.
  • Page 238 Every participating device must be able to perform at least one of the following roles; talker, listener or controller. The HP 3324A can be either a talker or a listener. The full flexibility and power of the HP-IB is realized when a controller is added to the system.
  • Page 239: Capabilities Of The Hp-Ib

    Address Capability Primary addresses: 31 talk, 31 listen; secondary (2-byte) addresses: 961 talk, 961 listen. 1 talker and 14 listeners, maximum, at one time. The HP 3324A has only primary address capability. Table D-l lists the talk and listen HP-IB addresses.
  • Page 240: Bus Structure

    This line is used by the active controller to define how information on the data lines should be interpreted by other devices on the bus. When ATN is low (true) the HP-IB is in the command and the data lines should be interpreted as bus mode...
  • Page 241: Srq . Service Request

    The HP 3324A responds to a serial poll by putting its status byte on the data lines (see Chapter 10 Status and Event Reporting).
  • Page 242 Talk/Listen Each HP-IB device has at least one talk and one listen address unless the device is either totally transparent or Addresses is a talk-only or listen-only device. Device addresses are used by the active controller in the command mode (ATN true) to specify the talker (via a talk address) and the listener(s) ( via listen addresses).
  • Page 243 Table D-l. HP-16 Addresses Device Binary Address Characters Address Address Talk Listen Space 0000 0000 0000 0001 0000 0011 0000 0010 0000 0100 0000 0101 0000 0110 0000 1000 0000 0111 0000 1001 0000 1010 0000 1011 0000 1100 0000 1101...
  • Page 244: Clear

    The HP-IB interface system operates in one of two modes, controlled by the ATN bus management line: command mode (ATN true) or data mode (ATN false). (if an HP controller is used, the bus management lines are configured automatically and all necessary command strings are issued).
  • Page 245: Clear Lockout/Set Local

    (front panel) control. If local lockout is not in effect, the HP 3324A responds by returning to front panel control. The Remote indicator on the front panel extinguishes if the HP 3324A is in Remote prior to the Local command. Local Lockout The local lockout command disables the m front panel key to avoid operator interference.
  • Page 246: Serial Poll

    Bits 0, 1, 2, and 3 in the status byte may initiate an SRQ depending on the setting of the status byte mask. The status byte may be masked to select which of the four bits cause the HP 3324A to issue the SRQ. Trigger...
  • Page 247: Hp 3324A Command Syntax

    HP 3324A Command Syntax Introduction This appendix contains the syntax of the HP-IB commands for the HP 3324A. The list of commands in this appendix are given in alphabetical order. For more information see Chapter 9 Remote Control of the HP 3324A and Chapter 11 HP-IB Commands.
  • Page 248: Hp-Ib Command List

    HP-IB Command The following table shows all of the HP-IB commands available, alphabetical order: List Table E-l. HP-IB Command List Description Description Mnemonic Phase Amplitude Calibration Recall State Amplitude Al’ Assign Zero Phase Rear or Front Signal Output H P 3325A Compatible...
  • Page 249 AC; Amplitude The AC command performs an amplitude calibration. If calibration is not successful, the FAIL bit of the status Calibration register is set. Command is available on the HP 3325A. Syntax HP 3324A Command Syntax E-3...
  • Page 250 The AM command sets the amplitude of the main signal. Sending IAM causes the instrument to output its current amplitude. Instrument Preset Value: 1.0 mV(p-p) Both commands are available on the HP 3325A. Syntax Table E-2. AM ‘cvalue” Range Given “units” “value” Range Description High Voltage 0.001 to 10.0...
  • Page 251 Zero AP; Assign The AP command assigns the current phase value to zero; subsequent changes in phase are referenced to this Phase point. Command is available on the HP 3325A. Syntax HP 3324A Command Syntax E-5...
  • Page 252 CM; HP 3325A The CM command switches the HP 3324A to the HP 3325A compatible mode or to the non-compatible mode Compatible at power-on. The difference between the two modes is explained in the following table: HP 3325A Compatible Mode...
  • Page 253 Table E-4. ICM Response Format Query Command Response Format HP 3324A Command Syntax E-7...
  • Page 254 IFR causes the instrument to output its current frequency. Instrument Preset value: 1 KHz Both commands are available on the HP 3325A. Syntax Table E-5. FR “value” Range Given “units” Table E-6. IFR Response Format...
  • Page 255 FU; Waveform The FU command selects the waveform function for the main signal output. Function Instrument Preset value: Both commands are available on the HP 3325A. Syntax Waveform Selections for “digit” Waveform DC only. Selects Selects Sinewave. Selects Squarewave. Selects Triangle.
  • Page 256 HV; High-Voltage The HV command controls the high-voltage amplifier option for the main signal output. output Instrument Preset value: 0 Command is available on the HP 3325A. Syntax Enable high-voltage amplifier. Table E-8. IHV Response Format Query Command Option Installed? Response Format...
  • Page 257 The query returns a one-digit Query code. If no error occurred, 0 is returned. Issuing the command clears the error code to 0. Command is available on the HP 3325A. Syntax Table E-9. IER Response Format Comman d Response Format...
  • Page 258 “0” or “00X” will be shown, where “0” means that the option is not installed and “00X” (X=option number) means that the option is installed. Command is not available on the HP 3325A. Syntax Table E-10. IOPT Response Format Response...
  • Page 259 The query returns a one-digit Query code. If no error occurred, 0 is returned. Issuing the command clears the error code to 0. Command is not available on the HP 3325A. Syntax Table E-11. ISE Response Format Query C ommand Response Format...
  • Page 260 RAM/ROM test failed (ROM=signature test) MFP (MC 68901) Test failed Device bus test failed Display test failed (display handshake) DAC test failed (analog voltage generation) FRAC-N chip test failed VCO test failed Sweep timer test failed Offset test, failed E-14 HP 3324A Command Syntax...
  • Page 261 HP 3324A program string has been accepted and all but the last character processed. In mode 2, the buffered mode, the execution control depends on the command group to which a command belongs.
  • Page 262 IXMF commands, respectively, when an index of 1 is used. To set and read the marker frequencies of other sweep intervals, use the XMF and IXMF commands. Instrument Preset value: 5.0 MHz Both commands are available on the HP 3325A. Syntax Table E-13. MF Range Given “units”...
  • Page 263 IMKS causes the instrument to output a list of the activated markers. Use the MMF command to set the frequency of the various markers. Instrument Preset value: All 9 markers Both commands are not available on the HP 3325A. Syntax Meaning 1 to 9 Table E-15. IMKS Response Format...
  • Page 264 Both commands are not available on the III’ 3325A. Syntax “index” I 1 to 9 1 Upto 9 separate markers are allowed. E-16. MMF Table “value” Range Given “units” Table E-17. IMMF Response Format Response Format Query Command I M M F E-18 HP 3324A Command Syntax...
  • Page 265 The MMS command selects whether multi-intervals or multi-markers is activated. Multi-Interval/ Marker Instrument Preset value: 1 (Multi-intervals) Both commands are not available on the HP 3325A. Meaning Multi-intervals selected, one marker/interval. Multi-markers selected, one interval. Table E-18. IMMS Response Format Query C ommand...
  • Page 266 (unmasked). They may be enabled or masked in any combination. Instrument Power-on,HP-IB Clear value: @ (no bits enabled) Instrument Preset value: not changed The MS command is available on the HP 3325h, but the IMS command is not. Syntax E-20 HP 3324A Command Syntax...
  • Page 267 Table E-19. Status Byte Mask Characters “character” System Fail Sweep Start Sweep Stop Program Em-o: O f f O f f O f f Table E-20. IMS Response Format Query Command Response Format HP 3324A Command Syntax E-21...
  • Page 268 The MT1 command is used to set the sweep time for the single interval when multi-markers are used. Sweep Time Instrument Power-on value: 1.0 sec. Both commands are not available on the HP 3325A. Syntax Table E-21. MTI “valuen Range Given “units” Description 100000 Seconds Table E-22.
  • Page 269 The MUP command is used to the single interval when multi-markers are used. Stop Frequency Instrument Power-on value: 10 MHz Both commands are not available on the HP 3325A. Syntax v a l u e Table E-23. MUP “value” Range Given “units” Range (sine) 0.0 to 21000000.000...
  • Page 270 The MUT command is used to set the start frequency MUT; Multi-Marker for the single interval when multi-markers are used. Start Frequency Instrument Power-on value: 1 MHz Both commands are not available on the HP 3325A. Syntax Table E-25. MUT “value” Range Given “units” “value” Range (sine) “units” Description 0.0 to 21000000.000...
  • Page 271 Chapter 5 under the heading “AC with DC Offset”. Instrument Preset value: 0.0 V(p-p) Both commands are available on the HP 3325A. Syntax v a l u e u n i t s Table E-27. OF 5ralue” Range Given “units”...
  • Page 272 OOF; Output The OOF command switches the main signal output on or off. Enable Instrument Preset value: 0 Both commands are not available on the HP 3325A. Syntax Meaning Disable main signal output. Enable main signal output. Table E-29. IOOF Response Format...
  • Page 273 PC; Phase The PC command causes a phase calibration between two HP 3324A’s, one of which has option 003 (slave) Calibration installed and the other option 004 (master). The PC command can only be sent to the instrument with option 003 installed.
  • Page 274 Both commands are available on the IIP 3325A. Syntax v a l u e 1 Table E-30. PH “value99 Range Given “units” “value” Range “units” Description -720.0 to 720.0 Degrees Table E-31. IPH Response Format Query Command Response Format E-28 HP 3324A Command Syntax...
  • Page 275 If either the SNR or SNI command is sent, all of the stores are cleared. This also happens with an instrument reset, and at power-on if the instrument is in the HP 3325A compatible mode. Command is available on the HP 3325A.
  • Page 276 OOF is the command used to control the main output signal, RF only determines where the signal will be present. 1 (front) Instrument Preset value: Both commands are available on the HP 3325A. Syntax Table E-32. IRF Response Format Query Comma4 Respons Format...
  • Page 277 3 ms logarithmic interval). Instrument Preset value: 0 (automatic) Both commands are not available on the HP 3325A. Syntax v a l u e Table E-33. RTT’ “value” Range Given “units” 100000 Seconds Table E-34.
  • Page 278 The SC command starts a continuous sweep. If the instrument is already sweeping, this command stops the Continuous Sweep sweep and does not restart it. FR can also be used to stop a sweep. Command is available on the HP 3325A. syntax E-32 HP 3324A Command Syntax...
  • Page 279 ISGS Intervals causes the instrument to output a list of the sequence. Both commands are not available on the HP 3324A. Syntax Meaning 1 to any of “SNI” The number of the interval.
  • Page 280 XSM and commands. IXSM Instrument Preset value: 1 Both commands are available on the HP 3325A. Syntax Table E-36. ISM Response Format Query Command Response Format E-34 HP 3324A Command Syntax...
  • Page 281 Note Sending the SNI command causes all of the stores to be cleared, and all of the intervals to be set to the default value. Both commands are not available on the HP 3325A. Syntax Meaning 1 to 50 Enable this number of intervals.
  • Page 282 Sending the SNR command causes all of the stores to be cleared, and all of the intervals to be set to the default value. Both commands are not available on the HP 3325~4. S y n t a x s p a c e...
  • Page 283 To set and read the stop frequency of other sweep intervals, use the XSP and IXSP commands. Instrument Preset value: 10 MHz Both commands are available on the HP 3325A. Syntax u n i t s v a l u e Table E-39.
  • Page 284 If either the SNR or SNI command is sent, all of the stores are cleared. This also happens with an instrument reset, and at power-on if the instrument is in the HP 3325A compatible mode. Command is available on the HP 3325A.
  • Page 285 Start Frequency. If the instrument is already in the sweep-reset state, this command starts a single sweep. If the instrument is sweeping, this command stops the sweep and does not restart it. Command is available on the HP 3325A. Syntax HP 3324A Command Syntax E-39...
  • Page 286 1 MHz Instrument Preset value: Syntax Table E-41. ST “value” Range Given “units” 21000000.000 0.0 to 0.0 to 21000.000000 0.0 to 21.000000 1 MH 1 Megahertz 1 Table E-42. IST Response Format Query Command Response Format E-40 HP 3324A Command Syntax...
  • Page 287 Cable Delay connecting the Phase Cal Out of the master (option 004) to the Phase Cal In of the slave (option 003). Instrument Preset value: 0 Command is not available on the HP 3325A. Syntax value un I ts Table E-43. TCD %aluen Range Given “units”...
  • Page 288 TE; Perform The TE command causes the instrument to perform a self-test. Self-test Command is available on the HP 3325A. syntax E-42 HP 3324A Command Syntax...
  • Page 289 XT1 and IXTI commands. Instrument Preset value: 1.0 sec. Both commands are not available on the HP 3325A. Syntax v a l u e Table E-45. TI “value” Range Given “units”...
  • Page 290 Cable Delay the device-under-test with the rear-signal outputs of the two HP 3324A’s containing options 003 and 004, when an automatic phase calibration is made. Instrument Preset value: 0 ps Command is not available on the HP 3325A.
  • Page 291 MF and IMF commands, respectively, when an index of 1 is used. Instrument Preset value: 5.0 MHz Both commands are not available on the HP 3325A. Syntax v a l u e u n i t s Meaning to any of “SNI”...
  • Page 292 1 is used. (Linear sweep mode) Instrument Preset value: 1 Both commands are not available on the HP 3325A. Syntax Meaning 1 to any of “SNI” Sets mode of interval. Selects Logarithmic sweep mode. Table E-51. IXSM Response Format...
  • Page 293 The XSP and IXSP commands are identical to the SP and ISP commands, respectively, when an index of 1 is used. 10 MHz Instrument Preset value: Both commands are not available on the HP 3325A. Syntax un its v a l u e index Meaning 1 to any of “SNI”...
  • Page 294 ST and IST commands, respectively, when an index of 1 is used. Instrument Preset value: 1 Both commands are not available on the HP 3325-4. Syntax 1 to any of “SNI” Sets start frequency of interval. Table E-54. XST Yralue” Range Given “units”...
  • Page 295 Instrument Preset value: 1.0 sec. are not available on the III’ 332511. S y n t a x I to auy of “SNI” Table E-56. XTI “value” Range Given “units” 0.0 to 100000 Seconds Table E-57. IXTI Response Format HP 3324A Command Syntax E-49...
  • Page 297 Error Messages Introduction This appendix contains a list of all the messages that can be obtained when the HP 3324A encounters an error. The following list contains all of the errors possible when operating the HP 3324A from the front panel.
  • Page 298 MESSAGE: Command: “HPIB addr.: xx” Cause: You are trying to change the HP-IB address, while the device is being addressed. Remedy: Un-address the HP 3324A via the controller before changing the HP-IB address. MESSAGE: Command: The frequency is either too small or too Cause: large for the current waveform.
  • Page 299 MESSAGE: HVamp/dBm incomp Command: “highvolt ON/off’ or (Ampl Cause: The amplitude cannot be in dBm for the high-voltage option. Enter the amplitude in Vrms or V(p-p). Remedy: index out range MESSAGE: “MARKER” “INTERVAL” Command: An index for a sweep interval has been Cause: entered which does not exist.
  • Page 300 log swp up only MESSAGE: Command: A log sweep has been programmed to be Cause: swept downwards. Re-enter the log sweep stop-frequency Remedy: to be higher than that of the start-frequency. no log swp <lHz MESSAGE: Command: A log sweep has been set with a Cause: start-frequency of <...
  • Page 301 For a DC and AC signal see the amplitude/offset table in chapter 5. MESSAGE: phase cal Wed Command: Cause: The attempted phase calibration has failed. Check to make sure that the two HP Remedy: 3324As are connected properly. MESSAGE: phase out range Command: (Phase) Cause: The phase value entered is not valid.
  • Page 302 MESSAGE: seq > 100 interv “INTERVAL” (m) Command: Cause: The sequence entered contains more than 100 intervals. Remedy: Reduce the number of intervals to 100 or less. MESSAGE: Command: [Reset Cause: The start frequency for the interval is too large. Remedy: Enter a smaller start frequency.
  • Page 303 MESSAGE: swp time < 10 ms Command: “SWEEP-TIME” Cause: The input sweep time is less than 10 ms. Remedy: Enter a sweep time for the interval of MESSAGE: swp time>lOOOOOs Command: “SWEEP-TIME” Cause: The input sweep time is greater than 100000 s.
  • Page 304 wrong seq index MESSAGE: Command: “INTERVAL” Cause: One of the indices in the sweep sequence does not exist. Remedy: Take the index out of the sweep sequence. F-8 Error Messages...
  • Page 305 The following lists the programming errors that can be obtained and evaluated using the IER program error query: Error Description Error Number Entry parameter out of bounds Invalid delimiter Frequency too high for waveform function Sweep time too small or too large Offset - amplitude incompatible Sweep frequency too large for function, start frequency too small,...
  • Page 306 The following lists the system errors that can be obtained and evaluated using the ISE system error query: System System Failure Description Failure Number Amplitude calibration failed Phase calibration failed External reference unlocked Main oscillator unlocked Self-test failed (report actual number to Service personnel).
  • Page 307 Backdating Introduction This appendix contains backdating information, which adapts this manual to instruments with serial numbers other than that shown on the title page. Changes are listed in the serial number order that they occured in the manufacture of the instrument. However, in adapting this manual to an instrument with serial number lower than that shown on the title page, apply the changes in reverse order.
  • Page 309 Bus management lines (HP-IB), D-4 Calibration amplitude, 5-8 Changing the bus address, 8-4, 9-11 Clear command, D-8 Clear lockout, D-9 Command groups, 9-7 Command syntax, E-l Compatibility mode setting the HP 3325A mode, 8-4 Compatibity mode HP 3325A compatible power-on state, 3-3 Connector Index-l...
  • Page 310 Data transfer modes, 9-7 DC offset, 5-10 high-voltage option, 5-13 no AC function, 5-11 with AC function, 5-11 Default address, 8-4, 9-11 Description of the HP-IB, D-l Display, 2-l Display indicators “compatib: yes/NO”, 8-4 2-5, 3-6 “HIGH-VOLT”, “highvolt: on/OFF”, 8-7...
  • Page 311 “HPIB addr.: xx”, 8-4 “INTERVAL”, 7-8 “LIN”, 7-9, 7-10 “LOG”, 7-9, 7-10 “M”, 3-11, 5-3 “MARKER”, 7-17 “OFFS”, 5-2 “PHASE”, 5-2 “phase cal MASTER”, 8-5 “phase cal SLAVE”, 8-5 “REAR-ONLY”, 2-3, 3-4 “REMOTE”, 9-4 “Reset”, 7-23 “ROM rev. xx.xx”, 8-7 “Run”, 7-23 “selft: std./all”, 8-5 “set default”, 8-l...
  • Page 312 frequency, function generator, modify mode, 3-13, 5-4 programming as a function generator, 1 I-10 programming as a multi-interval sweep generat)or, 1 l-l I programming as a multi-marker sweep generator, 11-13 recalling an instrument state, 3-16 saving an instrument state, 3-16 serial poll, 10-9 setting up a multi-interval sweep, 7-l 1 setting up a multi-marker sweep, 7-20...
  • Page 313 9-4 HP 3324A capabilities, 9-l logic levels, C-7 messages, 9-6 setting the address, 8-4, 9-11 Impedance of high-voltage output, 2-3, 2-4, A-11 of main signal output, 2-3, 2-4 Input signals, 3-4 Instrument introduction, l-l Interpreter, 9-2 Intervals setting number of, 8-3...
  • Page 314 7-22 3-11 3-11 2-1, 7-l 2-1, 7-l Linear sweep interval, 7-4 Local lockout, D-9 Local mode, l-2 Logarithmic sweep interval, 7-4 Main signal output connector, 2-3, 2-4, 3-4 Marker connector, 2-5 Marker frequency m u l t i - i n t e r v a l 7 - 5 , 7 - 1 5 sweep, Marker output connector, 3-7 Messages...
  • Page 315 offset, 5-10 Omitting the mnemonic, 9-10 Options available, B-l high-stability frequency reference, B-2 high-voltage output, B-3 phase calibration, B-4 Output signals, 3-4 Oven output connector (10 MHz), 2-4, 3-6 Parameter default values, 8-l Parameter keys, 2-2, 5-l Parameters amplitude, 5-7 DC offset, 5-10 frequency, 5-6 phase, 5-14...
  • Page 316 Setting the bus address, 8-4, 9-11 Setting up a multi-interval sequence, 7-8 Setting up a multi-marker interval, 7-17 Shift key, 3-9 Signals input/output, Single sweep, 7-7, 7-17 Software revision, 8-7 Specifications, A-l forcing an, 10-6 State sweep default, 7-2 turn-on, 3-2 Status byte, 10-l Status reporting, 10-l Stores...
  • Page 317 Talk/listen addresses, D-6 Tests self-test, 8-5 Trigger (HP-IB), D-10 Turn-on state, 3-2 Utilities, Viewing the bus address, 8-4, Y-11 Voltage AC line, C-3 Warm-up time, A-2 Waveforms, 6- 1 X-drive connector, 2-5, 3-8 Z-blank connector, 2-5, 3-7 Index-9...
  • Page 319 Sales and Service Office Directory Information Hewlett-Packard products are sold and supported worldwide through HP offices. To contact the closest HP Sales and Service Office, please check your telephone directory, or contact one of the HP Headquarters listed below. Asia Hewlett-Packard Asia Ltd.
  • Page 320 Eastern Europe Lieblgasse 1 P.O. Box 72 1222 Vienna Austria Telephone: (222) 2500-O Telex: 13 4425 HEPA A Hewlett-Packard S.A. Northern Europe V.D. Hooplaan 241 P.O. Box 999 118 LN 15 Amstelveen Netherlands Telephone: 20 547 9999 Telex: 189 19 hpner South East Europe Hewlet-Packard S.A.
  • Page 321 Middle East and Hewlett-Packard S.A. Middle East/Central Africa Sales H.Q. Central Africa 7, Rue du Bois-du-Lan P.O. Box 364 1217 Meyrin 1 Geneva Switzerland Telephone: (022) 83 12 12 Telex: 27835 hmea Cable: HEWPACKSA Geneve United Kingdom Hewlett-Packard Ltd. Nine Mile Ride Wokingham Berkshire, RGll 3LL Telephone: 0344 773 100...
  • Page 322 Midwestern USA Hewlett-Packard 5201 Tollview Drive Rolling Meadows IL 60008 Telephone (312) 255-9800 Southern USA Hewlett-Packard 2000 South Park Place Atlanta CA 30339 Telephone: (404) 955-1500 Western USA Hewlett-Packard 5161 Landershim Roulevard North Hollywood CA 91601 Telephone: (818) 505-5600 Hewlett-Packard Other International Intercontinental Operations...
  • Page 324 HEWLETT PACKARD Edition 1 E0489 Reorder Number: 03324-90011 Printed in FRG, 08/91, English...

Table of Contents