Ventilation
Requirements
The HP 8110A is fit,ted with two cooling fans. Make sure
that there is adequate
clearance of 3 inches (75 mm) at
the rear and l/2 inch (12 mm) at. t,he top and bottom
t,o
ensure adequate
airflow.
If the airflow is restricted
t,he
internal
operating
temperature
will be higher, reducing
the instrument's
reliability
or causing the instrument's
thermal-protection
circuits to automatically
switch off the
instrument.
Thermal
Protection
Overheating Detection
The HP 8110A monitors
its internal
temperature
in the
region of the power supply. If the temperature
exceeds
approximately
BOOC, the power supply is switched
off.
The inswument
will switch on again if the temperature
falls below approximately
77OC.
Fan Failure
If either of the fans is prevented
from operating
by a
blockage, or the power supply to the fans is interrupted,
the power supply is automatically
switched
off within
3
t,o 4 seconds. Note that after the fault condition
has been
fixed, the instrument
must remain switched
off for at
least 2 minutes
to allow the detection
circuit to recover.