<according to the Medical Safety/EMC standard EN 60601-1-2>
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information provided in the ACCOM-
PANYING DOCUMENTS.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
Electro Static Discharge (ESD)
1 Electro Static Discharge (ESD) warning symbol
2 The areas identifi ed with this symbol should not be touched as much as possible because
they are sensitive to ESD and may cause equipment malfunction and failure due to ESD.
When it is inevitable to touch these areas in installation, maintenance, or inspection, the
following ESD precautionary procedure should be used.
3 ESD precautionary procedure
(1) The power plug should be disconnected for safety's sake fi rst.
(2) The worker should wear an antistatic strap and attach it to the ground pin of the power
plug or the metal sheet on the bottom surface of the product. This eliminates the potential
difference between the product and the worker to prevent electro static discharge. (When
no antistatic strap is available, the worker may work while touching the ground pin of the
power plug or the metal sheet on the bottom surface of the product with one hand.)
(3) The necessary work for installation, maintenance, or inspection should be performed
under the above-mentioned conditions.
4 It is recommended that all staff involved in the work in which they may touch the areas
identifi ed with the ESD warning symbol should receive an explanation of the ESD warning
symbol and training on the ESD precautionary procedures.
5 Required minimum contents in the explanation and training on the ESD precautionary pro-
cedures
(1) Explanation on the principle of occurrence of static electricity and the ESD precaution-
ary procedures should be provided.
(2) Training on the ESD precautionary procedures using the actual product should be pro-
vided.
Technical description
List of all cables and maximum length of the cable and other ACCESSORIES
Maximum length
2 m
AC power cord
2 m
Video cable
1.5 m
S-video cable
PAPER/INK
RIBBON SET
WARNING:
The use of ACCESSORIES and cables other than those specifi ed, with the exception of cables
sold by the manufacturer of the Model CP31W as replacement parts for internal components,
may result in increased EMISSIONS or decreased IMMUNITY of the Model CP31W.
WARNING:
The Model CP31W should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use is necessary, the Model CP31W should be observed to verify normal
operation in the confi guration in which it will be used.
INSTRUCTIONS FOR MEDICAL USE
Reference page in this operation manual
This page, the previous pages for safety and page
53 for accessories
This page, the previous pages for safety
This page, the previous pages for safety
Page 19 and 54 for PAPER/INK RIBBON SET