Calculate3 Subsystem; Calculate3:Average Subtree - HP 53131A/132A 225 MHz Programming Manual

225 mhz universal counter
Table of Contents

Advertisement

NOTE
:CALCulate3:AVERage:ALL?
Query Response
Chapter 4 Command Reference
Introduction

:CALCulate3 Subsystem

This subsystem performs post-acquisition statistics computation and data transfer.
Not until :CALCulate3:LFILter:STATe is set to ON will any of the
:CALCulate3:LFILter settings be used.
Not until :CALCulate3:AVERage[:STATe] is set to ON will any of the
:CALCulate3:AVERage settings be used.
The statistics results are unaffected by post-processing invoked with
:CALC[1|2]:IMM.

:CALCulate3:AVERage Subtree

This subtree collects together the commands associated with the statistics
capabilities.
The statistics results combine successive measurements to produce a composite
result
Not until :CALCulate3:AVERage[:STATe] is set to ON will any of one
:CALCulate3:AVERage settings be used.
This query returns all four statistics (i.e., mean , standard deviation, maximum ,
and minimum ).
Statistics should be enabled (:CALC3:AVER[:STATe] ON) before attempting to
query results.
Numeric data transferred as ASCII bytes in <NR3> format.
The number of significant digits will range from 1 to 15, depending on the
measurement resolution.
Numbers are separated by commas. The ordering of numbers within the
response is mean, standard deviation, minimum,
and maximum.
If :CALC3:AVER[:STATe] is OFF, four comma-separated Not a Number
9.91E37 values are returned and error -221 is generated.
If no valid result exists, four comma-separated Not a Number 9.91E37
values are returned and error -230 is generated.
Programming Guide
4-19

Advertisement

Table of Contents
loading

This manual is also suitable for:

53131a53132a