Chapter 4 Command Reference
:CALCulate3 Subsystem
Query Response
Numeric data transferred as ASCII bytes in <NR3> format with eleven significant
digits.
Comments
:CALCulate3:LFILter:STATe <Boolean>
Sets or queries the statistics filter enable. When set to ON, only measurements
(scaled and offset if math is enabled) which are within the filter limits are
combined into the statistics processing. When set to OFF, all measurements,
whether they are within or without the filter limits are combined into the statistics
processing.
Query Response
Comments
*RST: OFF
Related
Stats
Front-Panel
Keys
:CALCulate3:LFILter:UPPer[:DATA] <numeric_value>
[HZ | S | DEG]
Sets or queries the statistics filter upper limit.
If limit filtering is enabled (:CALC3:LFIL:STAT ON), any measurements above
this value will not be combined into the statistics computation.
If math is enabled (:CALC:MATH:STATe ON), the limit value specified should
take into account that the filtering is on measurements that have been scaled and
offset.
4-24
*RST: 0.0000000000
This command couples :CALC2:LIM:LOW to the same value.
Updating the lower limit value causes the limit counts (:CALC2:LIM:FCO,
:CALC2:LIM:PCO) to be cleared.
The front panel menu item is not always able to display all of the significant
digits of this value. When this is the case, the displayed value is different from
the actual value in that the displayed value has been rounded. However, using
the front panel Enter key, while this value is in the 11-digit display, will
update the actual value to the displayed (rounded) value.
Single ASCII-encoded byte, 0 or 1.
A value of 0 indicates OFF; a value of 1 indicates ON.
Programming Guide