Entering Pressure Curve
Data
Pressure Curve Menu
[S/N]
CURVE
Probe Serial #
P: 123456
Pressure Curve Menu
S/N
[CURVE]
#
mA
Psig
1
+0.000 0.000
#
mA
Psig
1
+1.000 0.000
#
mA
Psig
1
+0.000 2.000
Repeat the three previous steps to enter the remaining data.You must
enter the necessary data for zero and span.
You may now do one of the following:
Enter data for a different channel or type of sensor by pressing the
[NO]
key until you reach the Programming Menu, and then refer to
Entering Calibration Data for New Probes/Sensors on page 3-39.
Refer to another section and perform a different procedure. Refer
to Figure 3-2 on page 3-48 and Figure 3-3 on page 3-49 to navigate
through the user program.
Exit by pressing the
IMPORTANT:
You must configure a channel AND enter the
necessary calibration data in order for a sensor to
work properly.
1
Use the arrow keys to move the
brackets to S/N and press
1
Enter the serial number and press
[YES].
Then press
the Curve Menu.
1
Use the arrow keys to move the
brackets to CURVE and press
[YES].
1
Making sure the cursor is
positioned below the # sign,
enter a 1 (for zero) or 2 (for
span). Press
the right arrow key.
1
Enter the mA value. Press
and press the right arrow key.
1
Enter the psig value. Press
[RUN]
key.
[YES]
.
[NO]
to return to
[YES]
and then press
[YES]
[YES].