HP 3300A Technical Manual page 17

Generator, signal sg-747/u
Table of Contents

Advertisement

Section V
Instrument Type
Electronic Counter
Distortion Analyzer
Oscilloscope
Probe 10:1
DC Voltmeter
Resistor
Resistor
Resistor
Capacitor
Variable Line Voltage
Transformer
DC Power Supply
AC Voltmeter
Printed Circuit
Extender Board
Printed Circuit
Extender Board
Table 5-1. Required Test Equipment
Required Characteristics
Range: dc to 100 kHz
Accuracy: 0.1%
Range: 10 Hz to 100kHz
Freq. Accuracy: +2%
Sensitivity: 0.3%fullscale
Input: 1 volt rms
Sensitivity: 50 mV/cm
Bandwidth: dc to 30 MHz
Bandwidth: dc to 30 MHz
Division Accuracy: ±2%
Accuracy: + 1% F. S.
Range: 10 mV to 50 V
Input Impedance: 10 MΩ
600 ohms
1/4 watt
+5%
50 ohms
1/4 watt
+5%
20 K
1/4 watt
+5%
1 µ F 50 V
Range: 100 to 130 V
Range: 0 - 10 volts, 500 mA
Range:10 Hz to 4 MHz
30 mV to 300 V
full scale
15 Pin
22 Pin
Use
Performance Checks,
Adjustment and
Calibration
Performance Checks
Performance Checks,
Adjustment
Calibration, Repair
Performance Checks,
Adjustment and
Calibration, Repair
Adjustment and
Calibration, Repair
Performance Checks
Performance Checks
Adjustment and
Calibration
Adjustment and
Calibration
Performance Checks
Performance Checks,
Adjustment and
Calibration
Adjustment and
Calibration
Repair
Repair
5-0
Model 3300A
Recommended Instrument
-hp- 5245L Electronic
Counter with 5262A Plug-
in Time Interval Unit
-hp- Model 331A Distortion
Analyzer
-hp- 175A Oscilloscope
with -hp- Plug-in 1750B
Vertical Amplifier
-hp- 10001A Probe 10:1
-hp- 3440A Digital Volt-
meter with Plug-in -hp-
Model 3443A
-hp- Part No. 0730-0010
-hp- Part No. 0683-5105
-hp- Part No. 0686-2035
-hp- Part No. 0160-0859
Superior Type UCIM
-hp- 723A Power Supply
-hp- 400F/FL Voltmeter
-hp- Part No. 5060-0049
-hp- Part No. 5060-0630