Square Wave Response; Sync Output; Remote Frequency Control Check; Remote Frequency Control Test Setup - HP 3300A Technical Manual

Generator, signal sg-747/u
Table of Contents

Advertisement

Section V
5-14.
Repeat 5-13 on CHANNEL B.
remain > 15 volts peak-to-peak.
5-15.
Repeat
5-13
instrument with the 50 ohm resistor. CHANNEL A and
CHANNEL B voltage output should be > 2 volts peak-to-
peak.
5-16.

SQUARE WAVE RESPONSE.

a.
Test equipment required: Oscilloscope (-hp-
Model 175A/1750B) and 10:1 Probe (-hp-
Model 10001A).
b.
Connect CHANNEL A OUTPUT without a
load to the oscilloscope using the 10:1
Probe, and set the 3300A controls as
follows:
CHANNEL A function .................SQUARE
FREQUENCY dial ......................10
RANGE switch............................X10K
c.
Verify: Rise and fall time
Sag
Overshoot (full output) <5%
Symmetry error
5-17
Repeat 5-16 on CHANNEL B.
Figure 5-1. 600 ohm or 50 ohm Load Output Test Setup
Limit should
and
5-14
except
load
<250 nano sec.
<1%
<1%
Figure 5-2. Remote Frequency Control Test Setup
5-18.

SYNC OUTPUT.

a.
the
b.
FREQUENCY dial ...................... 10
RANGE switch ........................... X1K
c.
5-19.

REMOTE FREQUENCY CONTROL CHECK.

a.
VOLTAGE APPLIED TO FREQUENCY
CONTROL BNC SHOULD BE LIMITED
TO A VALUE BETWEEN 0 AND
NEGATIVE 15 VOLTS.
OUTSIDE
DAMAGE THE INSTRUMENT.
5-2
Test equipment required: Oscilloscope (-hp-
Model 175A/1750B) and 10:1 Probe (-hp-
Model 10001A).
Connect SYNC OUTPUT to oscilloscope
and set 3300A controls as follows:
Pulse should be > 10 volts peak-to-peak
and < 5 microsecond duration.
Test equipment required: DC Power Supply
(-hp- Model 723A) and Oscilloscope (-hp-
Model 175A/1750B).
CAUTION
VOLTAGES
THIS
RANGE
Model 3300A
WILL

Hide quick links: