If You Have a Problem

"No Configuration File Loaded"

"No Configuration File Loaded"
This is usually caused by trying to load a configuration file for one type of
module or the system into a different type of module.
q
Verify that the appropriate module has been selected from the Load
{module} from File {filename} in the HP 16500A,B disk operation menu.
Selecting Load {All} will cause incorrect operation when loading most
preprocessor interface configuration files.
See Also
"To Load a Configuration File" in chapter 3, "File Management."

"Selected File is Incompatible"

This occurs when you try to load a configuration file for the wrong module.
Ensure that you are loading the appropriate configuration file for your logic
analyzer.

"Slow or Missing Clock"

q
This error message might occur if the logic analyzer cards are not firmly
seated in the HP 16500A/B or HP 16501A frame. Ensure that the cards
are firmly seated.
q
This error might occur if the target system is not running properly.
Ensure that the target system is on and operating properly.
q
If the error message persists, check that the logic analyzer pods are
connected to the proper connectors. See the User's Guide for your
preprocessor interface to determine the proper connections.
q
For HP 16510A Logic Analyzers, check the preprocessor interface power
fuse in the logic analyzer.
514