To Set Up A Group Run Of Modules Within The Hp 16500B - HP 16500B User Manual

Logic analysis system
Hide thumbs Also See for 16500B:
Table of Contents

Advertisement

To set up a group run of modules within the
HP 16500B
Modules are armed in the configuration tree by either an individual module or
the Group Run field. When armed, a module begins searching for the input
that will satisfy its trigger specification. To obtain a specification that triggers
on the arm signal, specify to trigger on "anystate."
Select the Intermodule menu.
1
Select the second field down from the top on the left, then select
2
Group Run.
Select the name of each module you want to include in the
3
measurement.
The modules are listed under "Modules" on the right side of the menu.
Select Group Run if you want to arm this module immediately when
a
the Group Run begins.
Select the name of another module if you want the other module to
b
arm the present one when the other module finds its trigger.
Select Independent if this module should not be included in the
c
intermodule measurement.
Select the Group Run field in the upper right hand corner.
4
The group run begins. The modules attached directly to the Group Run field
immediately begin searching for their respective trigger conditions. When a
module finds its trigger, it arms any modules attached to it in the Group Run
tree.
Intermodule Measurements

To set up a group run of modules within the HP 16500B