i.
REMOVE the sample holder. RETRIEVE your TEM grid. REINSERT the holder into the
CompuStage/column of the microscope.
j.
CHECK the Scheduler to determine if there is another user after you.
k.
If you are NOT the last user of the day, refill the LN
B. If you are the last user of the day:
a. REMOVE the LN
dewar, and PLACE the microscope dewar on the counter. PLACE a towel below the
dewar platform to collect condensing moisture.
b. On the CRT, GO to the Vacuum page, PRESS the Cryo key, SET the time for six hours
and PRESS the Start key.
C. Retrieving Images from the Support PC (SPC). Note: Files are deleted on the SPC every
three months.
a. COPY your images to a USB drive, LOG OFF the SPC and TURN OFF monitors.
b. CLOSE the AMTv600 program and LOG OUT of the PC.
c. TURN OFF the Panel light with the Panel Dim knob and CRT Screen with the Data Dim
knob.
D. Last Steps
a. SIGN the logbook and LEAVE any relevant comments. If you START the cryo cycle,
please INDICATE in the comments.
b. TURN OFF any lights.
c. EXIT the room and CLOSE the door.
anti-contaminator dewar, pour the remaining LN
2
dewar.
2
into the large
2
2017.07.07
20