c. On the PC screen, OPEN the CCD camera program by CLICKING the AMTV600 icon.
d. On the CRT, PRESS Ready, Mode, choose configuration page. Turn the filament
knob slowly until the Ext voltage reads 3.8. As a rule of thumb, rotate the knob
until you hear two clicks; wait; rotate again for two clicks; wait; repeat until
reaching 3.8.
e. OPEN the valve (from Close to Open,
counterclockwise) to the right of the
microscope. You should see the
electron beam on the fluorescent
screen. Move the region of interest of
your sample to the middle of the
viewing area with the Joystick.
f. On the TEM control panel, PRESS the
Auto Focus button.
g. On the AMTV600 window, CHANGE the
setting from Speedlive (button in red)
to Qualitylive by pressing the button
once.
2017.07.07
8