Parameters (HART)

The parameter groups are followed by the parameters within each group. The parameter
tables show the values you need to enter and are followed by additional information
when necessary. Factory settings are displayed after the parameter name, where
applicable.
Note: Parameter menus are dependent on selections made by the user.
In some cases, parameter menus are renumbered according to the
selection made at the previous menu level. See table below for an
example.
When User chooses Parameter 4.2.2. 1 :
Application Type = Liquids (Process)
2
4.2.2.
: Surface
3
4.2.2.
: Deadband
4
4.2.2.
: Correction Factor
5
4.2.2.
: Filling Speed
6
4.2.2.
: Reflectivity
7
4.2.2.
: Failsafe Level
8
4.2.2.
: Failsafe Timer
9
4.2.2.
: Range Extension
7ML19985FH04
SITRANS LR 400 – INSTRUCTION MANUAL
When User chooses Parameter 4.2.2. 1 :
Application Type = User tank1
(not available when 4.2.2. 1 = User tank1)
2
4.2.2.
: Deadband
3
4.2.2.
: Correction Factor
(not available when 4.2.2. 1 = User tank1)
(not available when 4.2.2. 1 = User tank1)
4
4.2.2.
: Failsafe Level
5
4.2.2.
: Failsafe Timer
6
4.2.2.
: Range Extension
Page 32