Multi-Stream Phosphate Monitor, Model 8242/195; Coils; Holding Registers - ABB 8241 Operating Instructions Manual

8240 series. modbus communication supplement
Table of Contents

Advertisement

...10 MODBUS COILS AND REGISTERS

10.4 Multi-stream Phosphate Monitor, Model 8242/195

10.4.1 Coils

Coils used on Multi-stream monitors are identical to those used on single-stream versions – see Section 10.3.1.

10.4.2 Holding Registers

In addition to the Holding Registers listed in Section 10.3.2 above, Multi-stream monitors use the following Holding Registers:
Input Number
Read/Write
100
R
101
R
102
R
103
R
104
R
105
R
106
R/W
107
R/W
108
R/W
109
R/W
110
R/W
111
R/W
112
R/W
113
R/W
114
R/W
115
R/W
116
R/W
117
R/W
118
R/W
119
R/W
120
R/W
121
R/W
122
R/W
123
R/W
124
R/W
125
R/W
126
R/W
127
R/W
128
R/W
129
R/W
130
R/W
131
R/W
132
R/W
133
R/W
134
R/W
135
R/W
136
R/W
20
Description
Phosphate Concentration,
Stream 1
Phosphate Concentration
Stream 2
Phosphate Concentration
Stream 3
Phosphate Concentration,
Stream 4
Phosphate Concentration
Stream 5
Phosphate Concentration
Stream 6
Stream 1 Decimal Point
Stream 2 Decimal Point
Stream 3 Decimal Point
Stream 4 Decimal Point
Stream 5 Decimal Point
Stream 6 Decimal Point
Alarm 1 Relay Setpoint
Alarm 1 Relay On/Off Status
Alarm 1 Relay Action
Alarm 2 Relay Setpoint
Alarm 2 Relay On/Off Status
Alarm 2 Relay Action
Alarm 3 Relay Setpoint
Alarm 3 Relay On/Off Status
Alarm 3 Relay Action
Alarm 4 Relay Setpoint
Alarm 4 Relay On/Off Status
Alarm 4 Relay Action
Alarm 5 Relay Setpoint
Alarm 5 Relay On/Off Status
Alarm 5 Relay Action
Alarm 6 Relay Setpoint
Alarm 6 Relay On/Off Status
Alarm 6 Relay Action
Relay Configuration
Current Output 4 Range
Current Output 5 Range
Current Output 6 Range
Current Output 1 Range
Current Output 2 Range
Current Output 3 Range
Response/Entry
Scaled 0 to 20.0 or 0 to 60.0 with decimal places
Scaled 0 to 20.0 or 0 to 60.0 with decimal places
Scaled 0 to 20.0 or 0 to 60.0 with decimal places
Scaled 0 to 20.0 or 0 to 60.0 with decimal places
Scaled 0 to 20.0 or 0 to 60.0 with decimal places
Scaled 0 to 20.0 or 0 to 60.0 with decimal places
0 = 1 d.p., 1 = 2 d.p.
0 = 1 d.p., 1 = 2 d.p.
0 = 1 d.p., 1 = 2 d.p.
0 = 1 d.p., 1 = 2 d.p.
0 = 1 d.p., 1 = 2 d.p.
0 = 1 d.p., 1 = 2 d.p.
Scaled 0 to 20.0 or 0 to 60.0
0=Off, 1=On
0=Low, 1=High
Scaled 0 to 20.0 or 0 to 60.0
0=Off, 1=On
0=Low, 1=High
Scaled 0 to 20.0 or 0 to 60.0
0=Off, 1=On
0=Low, 1=High
Scaled 0 to 20.0 or 0 to 60.0
0=Off, 1=On
0=Low, 1=High
Scaled 0 to 20.0 or 0 to 60.0
0=Off, 1=On
0=Low, 1=High
Scaled 0 to 20.0 or 0 to 60.0
0=Off, 1=On
0=Low, 1=High
0=Concentration, 1=Out of Sample
Scaled 0 to 20.0 or 0 to 60.0
Scaled 0 to 20.0 or 0 to 60.0
Scaled 0 to 20.0 or 0 to 60.0
Scaled 0 to 20.0 or 0 to 60.0
Scaled 0 to 20.0 or 0 to 60.0
Scaled 0 to 20.0 or 0 to 60.0