Ma Step - GE Druck DPI 610 User Manual

Intrinsically safe portable pressure calibrator
Hide thumbs Also See for Druck DPI 610:
Table of Contents

Advertisement

Advanced Task

mA Step

To select one of the electrical output programs, press the OUTPUT key and proceed as
follows:
On selection of (e.g.) Linear, the output display window changes to show the selected
program of output currents:
Connect an external power source as shown on page 24.
Press RUN (F1) to run program. A flashing status display CHECK LOOP indicates a
fault in the external loop i.e. supply fault or open circuit.
Note: The dwell time at each step is approximately 10 seconds.
Press STOP (F1) when running to stop at any point. Press RUN (F1) to resume.
25
Select Output
K0430 Issue No. 2