Memory Operations

Downloading Procedure Files (DPI 615 instruments only)

THE RS232 INTERFACE MUST ONLY BE USED IN A SAFE AREA
Complete test procedures may be downloaded from a PC to the DPI 615 instrument via
the RS232 port. A procedure consists of a number of Druck Universal Command
Interface (DUCI) commands that are usually assembled by a linking management
software application (e.g.) Druck Intecal.
Before downloading a procedure, the instrument must be in the REMOTE mode. To place
the instrument into REMOTE mode, proceed as follows.
Connect the instrument's RS232 port to a free COM port on a PC.
Make sure that the COMMS parameters of the PC match those of the instrument
(refer to page 38).
Make sure that the instrument is not already running a procedure. If it is, quit the
procedure.
Download the procedure. Procedures are stored in the Data Log directory.
The following sequence shows a typical download sequence that starts with the
instrument in LOCAL mode.
After the procedure file has been downloaded, the instrument is usually returned to the
LOCAL mode by the last command in the procedure file. If the instrument remains in
REMOTE mode, switch it OFF and ON to reset it.