Calibrate External Sensors - GE Druck DPI 610 User Manual

Intrinsically safe portable pressure calibrator
Hide thumbs Also See for Druck DPI 610:
Table of Contents

Advertisement

Calibration

Calibrate External Sensors

Use the following procedure for calibrating external pressure sensors.
(1)
Connect the required external transducer to the EXT TRANSDUCER socket located
on the rear of the instrument.
(2)
Allow the instrument's temperature and the temperature of the external
transducer to stabilize in the calibration environment for a minimum of 1 hour.
(3)
Switch the instrument on, enter CALIBRATION mode and select EXT SENSORS
from the CALIBRATION menu:
(4)
Select the transducer to be calibrated from the transducer menu using the
cursor keys and press ENTER:
If the sensor to be calibrated is not in the directory, or no sensors are listed in the
directory, press the F1 key (ADD NEW SENSOR). This will place the sensor in the
directory, allowing it to then be selected.
(5)
Connect the pressure standard to the inlet of the external transducer, apply the
zero point pressure and store the zero point:
INT RANGES
EXT SENSORS
CHANGE PIN
55
K0430 Issue No. 2