GE Druck DPI 620 Genii User Manual page 160

Advanced modular calibrator
Hide thumbs Also See for Druck DPI 620 Genii:
Table of Contents

Advertisement

User Manual
4. Check the calibration is correct:
• Select the Resistance (source) function (Ref: Section
2.11).
• Apply the resistance values that are applicable to the
calibration (Ref: Table 8-13).
• Check the error is within limits (Ref: Table 8-13).
Table 8-13 Resistance (source) error limits
Ohms
(Ω)
0
100
200
300
400
1000
2000
3000
4000
06 June 2013
Excitation
uncertainty
(mA)
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
K0541 issue 1
Calibrator
(Ω)
0.0014
0.0016
0.0021
0.0028
0.0035
0.008
0.016
0.024
0.032
Druck DPI620 Genii
Permitted
DPI 620
error (Ω)
0.014
0.038
0.062
0.086
0.11
0.31
0.55
0.86
1.1
Page 160 of 175