GE Druck DPI 620 Genii User Manual page 58

Advanced modular calibrator
Hide thumbs Also See for Druck DPI 620 Genii:
Table of Contents

Advertisement

User Manual
Figure 2-23 CH1 Thermocouple Source CH2 Switch Test
Figure 2-23 shows a thermocouple switch test.
1. Set the applicable software options:
• The TC is set to source a temperature
• The UTILITY is set to SWITCH TEST. The AUTOMATION is
set to RAMP
2. Complete the electrical connections.
3. It is a CH1 function, the switch connections must be on
CH2.
4. For the Ramp process, set START and STOP values that
are applicable to the switch value.
5. To get an accurate switch value, set a long TRAVEL
period.
06 June 2013
K0541 issue 1
Druck DPI620 Genii
Page 58 of 175