Fault Reset; Device Calibration - Siemens Sitrans PROBE LU Instruction Manual

Profibus pa
Hide thumbs Also See for Sitrans PROBE LU:
Table of Contents

Advertisement

Fault Reset

Used to reset a fault message after an active fault has occurred and been corrected. A
manual reset is required only for certain faults, identified by an asterisk (*) in the General
Fault Code list on page 30.
To reset, key in the value of the fault code in question.
1.
Go to Reset Fault:
I
1.
dentification Menu
1. 2. Configuration
2.
Press the Right ARROW key
3.
Key in the number of the Fault Code to reset the status to normal operation.
4.
Press the Right ARROW key
Navigation mode.

Device calibration

Only four settings are required for a Quick Setup: High Calibration Point, Low Calibration
Point, High Level Point, and Low Level Point. (For an illustration, see
20.)
2.4. Sensor Calibration:
2.4.1. Sensor Units (default meters)
2.4.2. Calibration Type.
Values
2.4.3. Low Calibration Pt.
2.4.4. High Calibration Pt.
2.4.5. Unit [Level] (default percent)
2.4.6. Low Level Point
2.4.7. High Level Point
1.
Go to Calibration Type and verify that Dry Calibration is selected.
2.
Go to Low Calibration Point and enter the new value (default units are meters).
3.
Go to High Calibration Point and enter the new value (default units are meters).
4.
Go to Low Level Point and enter the corresponding value in percent (default is 0).
5.
Go to High Level Point and enter the corresponding value in percent (default is 100).
6.
SITRANS Probe LU is now ready to operate.
7ML19985JB02
SITRANS Probe LU (PROFIBUS PA) – INSTRUCTION MANUAL
1.2.4. Reset Fault
to open Edit mode.
again, to accept the change and return to
0
*
Dry calibration (preferred)
1
Wet calibration
Calibration
on page
Page 47