Setting Up a Pulse Signal

Task
Set up a continuous pulse signal with 50 ns period, a pulse width of 30 ns,
a leading edge of 6 ns, a trailing edge of 10 ns, an amplitude of 3.3 V and
an offset of 1.65 V (high level 3.3 V, low level 0.0 V).
0.0 V
Instructions
To set the operating mode and trigger mode as required:
1 Reset all parameters and modes by pressing SHIFT + STORE + 0.
2 Press the MODE/TRG softkey to enter the Mode/Trigger screen.
3.3 V
50 %
The default settings can be used for this example. For reference the
Mode/Trigger screen is displayed. There is nothing to change.
CONTINUOUS
PULSES
Single-Pulses at Out1
Pulse-Period: internal Osc
MODE/TRG
Pulse Width = 30 ns
Period = 50 ns
OUTPUT
LIMITS
Getting Started
Setting Up a Pulse Signal
MODIFY
*PulseStrm
Burst
TRG-LEV