Setting Up A Burst Signal - HP 81101A Quick Start Manual

Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

Setting Up a Burst Signal

Set up a burst signal with a burst repetition of 2 µs. Each burst consists of
Task
two double-pulses at a period of 500 ns. The pulse width is 100 ns, the
delay between the two pulses of a double-pulse is 200 ns. The levels are
2 Vpp amplitude and 0 V offset.
–1 V
Instructions
To set the operating mode and trigger mode as required:
1 Reset all parameters and modes by pressing SHIFT + STORE + 0.
2 Press the MODE/TRG softkey to enter the Mode/Trigger screen.
3 Move to CONTINUOUS and turn the knob to select T
4 Move to PULSES and select B
5 Move down and select D
6 Move left and select a burst length of 2.
7 Move down and select T
Double Pulse Delay = 200 ns
1 V
0 V
Pulse Period = 500 ns
The internal PLL is used to trigger the startable internal oscillator.
The PLL defines the burst repetition, the oscillator generates the
pulse period. This requires the following settings:
(burst repetition).
Burst Repetition = 2 µs
.
URST
(D
-P
OUBLE
OUBLE
ULSES AT
: PLL and set the period to 2.000 µs
'
RG
D BY
Getting Started
Setting Up a Burst Signal
.
RIGGERED
O
1).
UT
39

Advertisement