6. Configure the preset names for the Operators, if necessary. Refer to
Names (p.
15-19).

RRA Confirmation Screen

The RRA Confirmation Screen selection controls the presence of the confirmation screen displayed on
the viewer when the exam is closed. This screen provides a summary of the RRA classified images in the
exam.
Follow this process to enable or disable RRA Confirmation Screen:
1. From the Worklist screen, click [UTILITIES].
2. Select Preferences > Repeat/Reject Analysis.
3. Click RRA Confirmation Screen [ON] to enable the function.
 RRA Confirmation Screen [OFF] disables the function.
4. Click [CLOSE].
Export RRA Data and Report
The system stores RRA data for the last 100 days. This data may be exported in a report to a disk or USB
device.
IMPORTANT! RRA data is not saved with system backups. RRA data should be exported prior to per-
forming any software updates to the system.
IMPORTANT! RRA Data cannot be exported on DVD. Only CDs and USB devices can be used for RRA
export.
The exported data includes a summary report in html format
sheet
(Figure
13-7) with acquisition and classification details for all repeated and rejected images on the
system. The spreadsheet also contains JPEG links for all repeated and rejected images on the system.
The Excel spreadsheet includes the following information for each repeated/rejected image:
System ID
Automatic Protocol Code
Acquisition Date
Acquisition Time
Processing Date
Processing Time
RRA Classification
Repeat Reason
Operator Name
IMPORTANT! The RRA export data does not contain patient names: however, patient ID numbers, image
JPEGs, and accession number data are present. Proper HIPAA or privacy and internal or
local confidentiality standards should be practiced.
5495975-1EN Rev.9
© 2013-2017 General Electric Company. All rights reserved.
Anatomy
View
Acquisition Type
Link to JPEG Image
Patient ID
Accession Number
Size Selection
kVp
mAs
Chapter 13: Advanced Applications
Chapter 15: Preferences-Preset
(Figure
13-6) and a Microsoft excel spread-
mA
SID (mm)
Grid Status
Filtration (mm Cu)
Manual/AEC
Ion Chambers Selected
DAP
DEI
DEI Limits
13-9

Advertisement

Table of Contents
loading

Table of Contents