GE Optima XR646 Operator's Manual page 369

Digital radiographic system
Table of Contents

Advertisement

Figure 14-14 Failed QAP results
13. The system prompts you if the QAP test has failed and indicates which portion was affected by dis-
playing messages. The following list provides information to correct the problem.
 If the flat-field tests fail, check the collimator blade position and make sure they are fully open, i.e.
the collimator blades are not in the field of view.
 If the MTF phantom test fails, a message appears (Figure 14-15).
Figure 14-15 QAP Phantom not found message
 Check that the flat-field phantom is not in the collimator rail.
 Check that the tube and detector are vertically aligned.
 Check that the MTF phantom is inserted completely into the grid holder.
5495975-1EN Rev.9
© 2013-2017 General Electric Company. All rights reserved.
Chapter 14: Quality Assurance and Maintenance
14-12