GE ApexPro Service Manual page 161

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

External surfaces
2001989-203C
Precautions
Clean the exterior surface of the ATS on a regular basis in compliance with your
institution's infection control unit and biomedical engineering department.
Use the following precautions to prevent damaging the equipment:
WARNING
Disconnect AC-powered equipment from the power line before
cleaning or disinfecting its surface. Turn off the power to battery-
powered equipment before cleaning or disinfecting its surface.
WARNING
Never immerse the device in any liquid.
WARNING
Do not pour or spray any liquid directly on the device or permit fluid
to seep into connections or openings of the ATS. An electrical shock
hazard exists and/or damage may result if water is allowed to get
inside the ATS. Failure to follow these instructions could result in
equipment damage and/or electrical shock.
WARNING
Never use conductive solutions, solutions that contain chlorides,
wax, or wax compounds to clean the device.
WARNING
Never use solutions or products that contain the following:
Any type of Ammonium Chloride such as, but not limited to:
Dimethyl Benzyl Ammonium Chloride
Quaternary Ammonium Chloride solutions
Abrasive cleaners or solvents of any kind
Acetone
Ketone
Betadine
Alcohol-based cleaning agents
Sodium salts
WARNING
Never autoclave or steam clean the device.