8. Insert tubing into the pump assembly.
a. Extend the new tube to its respective pump and thread the end into the hole in the
b. Push the thumb tab to the extreme left, as before, and push the tube through until
c. Release the pump arm.
NOTE:
If you encounter difficulty in threading the tubing through the pump housing, lift the
operating lever to
pump housing to rotate slightly and relieve the interference.
Figure 6-11. Inserting Tubing into Pump Assembly
9. Connect the tube to its proper recessed nipple.
NOTE:
If difficulty is encountered in connecting a tube to its nipple, use forceps or a hemostat to
grasp and attach the tube.
CAUTION
Take care to not damage the tubing when using forceps or hemostats.
10. Replace the cannulas into their respective reagent bottles.
Page 6—10, Section 6: Minor Replacements and Adjustments
pump arm.
the plastic cuff is flush against the pump arm.
for just a few seconds. This will cause the rollers inside the
PRIME