General Troubleshooting; Figure 28 S/5 Am, Ccm General Troubleshooting Flowchart - GE Datex-Ohmeda S/5 Technical Reference Manual

Anesthesia monitor, critical care monitor
Table of Contents

Advertisement

6

General troubleshooting

Monitor not functioning
On/Stby sw itch "ON"?
Display connected and
pow ered? Keyboard
connected?
Yes
Mains
pow er cord connected?
Possible intrconnecting
cables connected?
No
Connect pow er cord and
interconnecting cables
Remove all modules and turn
pow er on
Stby-led lit?
Yes
Start-up texts
appear on
screen?
Yes
Clock etc. data
appears on the
screen?
Yes
Plug in a module
Module data
appears on the
screen?
No
Another
moudle w orks in
same place?
Figure 28
No
No
Yes
Monitor starts
functioning
No
No
Keyboard trouble. See troubleshooting in
No
No
No
See troubleshooting in Frame slot.
Yes
Yes
See troubleshooting in module slot.
Faulty Module Frame or Central Unit.
No
See troubeshooting in Frame slot.
S/5 AM, CCM general troubleshooting flowchart
Connect and check
Monitor starts
functioning?
Disconnect and reconnect the
pow er cord.
Monitor starts
functioning?
Yes
Command Bars slot.
Faulty Central Unit.
Pick up the next module.
Faulty module.
General troubleshooting
The problem w as caused by
Yes
loosen unit(s)
The pow er supply unit w as shut
Yes
dow n by instant over voltage in the
mains.
The problem w as caused by
loosen pow er cord
Document no. M1137263-04