Figure 18 S/5 Compact Monitor General Troubleshooting Flowchart - GE Datex-Ohmeda S/5 Series Technical Reference Manual

Compact anesthesia monitor, compact critical care monitor
Table of Contents

Advertisement

General troubleshooting
6
General troubleshooting
Monitor not
functioning
Power cord
connected?
No
Connect power
cord
Yes
Monitor can be
Batteries were
Yes
started now?
empty
No
Possible AC/DC
On/Standby LED
unit failure.
Monitor can be
No
is ON or blinks?
Replace AC/DC
started now?
unit.
Yes
Yes
Frame Unit fan is
Display Unit
AC/DC Unit
Yes
running?
failure.
failure.
No
PMB failure.
No
Replace PMB.
Figure 18
S/5 Compact Monitor general troubleshooting flowchart
65
Document no. M1144951-004