Monitor Troubleshooting - GE LOGIQ P5 Service Manual

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
7-7-6

Monitor Troubleshooting

NOTE:
Reset the monitor settings to the factory defaults prior to troubleshooting the monitor.
Comments to remember about the
NOTICE
The monitor is the diagnostic monitor and is connected to AC 110/220 VAC.
The monitor powers down with the console.
Power output from AC Power assy via AC distributor in AC Transbox assy:
7-30
5
15"
monitor.
When the monitor is not connected to the AC power there is a leakage current which
when measured with a DVM can indicate high voltage.
When a load is applied, the voltage collapses to low voltage if the monitor is not
switched on.
Section 7-7 - Trouble Shooting Tree
LOGIQ™ P5 S
M
EVICE
ANUAL